Comparing blood pressure measurements between a photoplethysmography-based and a standard cuff-based manometry device

Dean Nachman, Yftach Gepner, Nir Goldstein, Eli Kabakov, Arik Ben Ishay, Romi Littman, Yuval Azmon, Eli Jaffe, Arik Eisenkraft

Open full text 107 citations

Abstract

Repeated blood pressure (BP) measurements allow better control of hypertension. Current measurements rely on cuff-based devices. The aim of the present study was to compare BP measurements using a novel cuff-less photoplethysmography-based device to a standard sphygmomanometer device. Males and females were recruited from within the general population who arrived at a public BP screening station. One to two measurements were taken from each using a sphygmomanometer-based and the photoplethysmography-based devices. Devices were considered equal if the mean difference between paired measurements was below 5 mmHg and the Standard Deviation (SD) was no greater than 8 mmHg. Agreement and reliability analyses were also performed. 1057 subjects were included in the study analysis. There were no adverse events during the study. The mean (± SD) difference between paired measurements for all subjects was -0.1 ± 3.6 mmHg for the systolic and 0.0 ± 3.5 mmHg for the diastolic readings. We found 96.31% agreement in identifying hypertension and an Interclass Correlation Coefficient of 0.99 and 0.97 for systolic and diastolic measurements, respectively. The photoplethysmography-based device was found similar to the gold-standard sphygmomanometer-based device with high agreement and reliability levels. The device might enable a reliable, more convenient method for repeated BP monitoring.

Open-access reader

About this research paper

What this paper is about

Repeated blood pressure (BP) measurements allow better control of hypertension. Current measurements rely on cuff-based devices. The aim of the present study was to compare BP measurements using a novel cuff-less photoplethysmography-based device to a standard sphygmomanometer device. Males and females were recruited from within the general population who arrived at a public BP screening station. One to two measurements were taken from each using a sphygmomanometer-based and the photoplethysmography-based devices. Devices were considered equal if the mean difference between paired measurements was below 5 mmHg and the Standard Deviation (SD) was no greater than 8 mmHg. Agreement and reliability analyses were also performed. 1057 subjects were included in the study analysis. There were no adverse events during the study. The mean (± SD) difference between paired measurements for all subjects was -0.1 ± 3.6 mmHg for the systolic and 0.0 ± 3.5 mmHg for the diastolic readings. We found 96.31% agreement in identifying hypertension and an Interclass Correlation Coefficient of 0.99 and 0.97 for systolic and diastolic measurements, respectively. The photoplethysmography-based device was found similar to the gold-standard sphygmomanometer-based device with high agreement and reliability levels. The device might enable a reliable, more convenient method for repeated BP monitoring.

Why it matters

OpenAlex reports 107 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Repeated blood pressure (BP) measurements allow better control of hypertension. Current measurements rely on cuff-based devices. The aim of the present study was to compare BP measurements using a novel cuff-less photoplethysmography-based device to a standard sphygmomanometer device. Males and females were recruited from within the general population who arrived at a public BP screening station. One to two measurements were taken from each using a sphygmomanometer-based and the photoplethysmography-based devices. Devices were considered equal if the mean difference between paired measurements was below 5 mmHg and the Standard Deviation (SD) was no greater than 8 mmHg. Agreement and reliability analyses were also performed. 1057 subjects were included in the study analysis. There were no adverse events during the study. The mean (± SD) difference between paired measurements for all subjects was -0.1 ± 3.6 mmHg for the systolic and 0.0 ± 3.5 mmHg for the diastolic readings. We found 96.31% agreement in identifying hypertension and an Interclass Correlation Coefficient of 0.99 and 0.97 for systolic and diastolic measurements, respectively. The photoplethysmography-based device was found similar to the gold-standard sphygmomanometer-based device with high agreement and reliability levels. The device might enable a reliable, more convenient method for repeated BP monitoring.

Key concepts: Photoplethysmogram, Sphygmomanometer, Cuff, Medicine, Intraclass correlation, Blood pressure, Gold standard (test), Diastole