2020•MetrologiaRequires access

A comparison of different calibration strategies for the trace analysis of soil samples by inductively coupled plasma mass spectrometry (ICP-MS)

Angelique Botha

Open publisher page 3 citations

Abstract

Abstract The National Metrology Institute of South Africa (NMISA) is expanding its capabilities in the field of the analysis of trace and ultra-trace levels of toxic and minor elements in environmental samples, with a special focus on soils to support projects of the University of Pretoria (UP) in the field of veterinary geology for the One Health Platform. The most accurate calibration strategy developed by the laboratory involves double isotope dilution analysis, where a stable isotope standard of the element of interest is used as the ideal internal standard. External calibration is the most commonly used calibration strategy for elemental analysis. To reduce the impact of instrumental drift on the accuracy of the measurement results, nominally the same amount of an internal standard is added to the samples and the standards. Standard addition is a calibration strategy that could eliminate the issue of matrix effects for trace and ultra-trace elemental analysis, because the calibration standards are added as spikes to aliquots of the sample to prepare the calibration curve. The purpose of the study was to compare the results obtained for the analysis of trace elements in soil samples using the calibration strategies of isotope dilution analysis; gravimetric external calibration with gravimetric internal standardisation; as well as gravimetric standard addition with gravimetric internal standardisation. The aim was to show that the application of gravimetric preparation to the more routine calibration approaches of external calibration and standard addition could achieve very comparable results and associated measurement uncertainties to isotope dilution analysis. The application of gravimetric sample preparation could assist routine laboratories, that have significant resource constraints, to improve the accuracy of their measurement results. These approaches could also be used in the experimental homogeneity and stability studies required for the production of reference materials, where high precision methods are required.

About this research paper

What this paper is about

Abstract The National Metrology Institute of South Africa (NMISA) is expanding its capabilities in the field of the analysis of trace and ultra-trace levels of toxic and minor elements in environmental samples, with a special focus on soils to support projects of the University of Pretoria (UP) in the field of veterinary geology for the One Health Platform. The most accurate calibration strategy developed by the laboratory involves double isotope dilution analysis, where a stable isotope standard of the element of interest is used as the ideal internal standard. External calibration is the most commonly used calibration strategy for elemental analysis. To reduce the impact of instrumental drift on the accuracy of the measurement results, nominally the same amount of an internal standard is added to the samples and the standards. Standard addition is a calibration strategy that could eliminate the issue of matrix effects for trace and ultra-trace elemental analysis, because the calibration standards are added as spikes to aliquots of the sample to prepare the calibration curve. The purpose of the study was to compare the results obtained for the analysis of trace elements in soil samples using the calibration strategies of isotope dilution analysis; gravimetric external calibration with gravimetric internal standardisation; as well as gravimetric standard addition with gravimetric internal standardisation. The aim was to show that the application of gravimetric preparation to the more routine calibration approaches of external calibration and standard addition could achieve very comparable results and associated measurement uncertainties to isotope dilution analysis. The application of gravimetric sample preparation could assist routine laboratories, that have significant resource constraints, to improve the accuracy of their measurement results. These approaches could also be used in the experimental homogeneity and stability studies required for the production of reference materials, where high precision methods are required.

Why it matters

OpenAlex reports 3 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Abstract The National Metrology Institute of South Africa (NMISA) is expanding its capabilities in the field of the analysis of trace and ultra-trace levels of toxic and minor elements in environmental samples, with a special focus on soils to support projects of the University of Pretoria (UP) in the field of veterinary geology for the One Health Platform. The most accurate calibration strategy developed by the laboratory involves double isotope dilution analysis, where a stable isotope standard of the element of interest is used as the ideal internal standard. External calibration is the most commonly used calibration strategy for elemental analysis. To reduce the impact of instrumental drift on the accuracy of the measurement results, nominally the same amount of an internal standard is added to the samples and the standards. Standard addition is a calibration strategy that could eliminate the issue of matrix effects for trace and ultra-trace elemental analysis, because the calibration standards are added as spikes to aliquots of the sample to prepare the calibration curve. The purpose of the study was to compare the results obtained for the analysis of trace elements in soil samples using the calibration strategies of isotope dilution analysis; gravimetric external calibration with gravimetric internal standardisation; as well as gravimetric standard addition with gravimetric internal standardisation. The aim was to show that the application of gravimetric preparation to the more routine calibration approaches of external calibration and standard addition could achieve very comparable results and associated measurement uncertainties to isotope dilution analysis. The application of gravimetric sample preparation could assist routine laboratories, that have significant resource constraints, to improve the accuracy of their measurement results. These approaches could also be used in the experimental homogeneity and stability studies required for the production of reference materials, where high precision methods are required.

Key concepts: Inductively coupled plasma mass spectrometry, TRACE (psycholinguistics), Calibration, Mass spectrometry, Analytical Chemistry (journal), Inductively coupled plasma, Environmental science, Chemistry

Related papers

Back to paper searchBrowse research topicsOriginal source
A comparison of different calibration strategies for the trace analysis of soil samples by inductively coupled plasma mass spectrometry (ICP-MS) — Research Paper | ScholarLens