The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging

Yun‐e Zhao, Ding Chen, Giacomo Savini, Qing Wang, Hongfang Zhang, Yili Jin, Benhao Song, Rui Ning, Jinhai Huang, Chenyang Mei

Open full text 22 citations

Abstract

Abstract Purpose To assess the repeatability and reproducibility of swept-source optical coherence tomography (SS-OCT) and Scheimpflug system and evaluate the agreement between the two systems in measuring multiple corneal regions in children. Methods Pachymetric and keratometric maps for both systems were evaluated. Central, midperipheral and peripheral corneal thickness (CT), keratometry and astigmatism power vectors were recorded. The three outcomes yielded by the same observer were used to assess intraobserver repeatability. The differences in the mean values provided by each observer were used to evaluate interobserver reproducibility. Within-subject standard deviation, test-retest repeatability (TRT) and coefficient of variation (CoV) were used to analyze the intraobserver repeatability and interobserver reproducibility. Paired T-test and Bland-Altman were used to appraise interdevice agreement. Results Seventy-eight eyes of 78 children were included. The CoV was ≤2.12 and 1.10%, respectively, for repeatability and reproducibility. TRT and CoV were lower for central and paracentral CT measurements than for peripheral measurements. The SS-OCT device generated higher precision when acquiring CT data, whereas Scheimpflug system showed higher reliability when measuring corneal keratometry. Although the CT readings measured using SS-OCT were significantly thinner than Scheimpflug device (P < 0.001), the central and thinnest CT values were still of high agreement. The interdevice agreement of keratometry measurement was high for the central corneal region and moderate for the paracentral and peripheral areas. Conclusions The precision of CT measurements by SS-OCT was higher, while the reliability of keratometry measurements by the Scheimpflug system was higher in children. Apart from the measured values in the central corneal region, the thickness and keratometry readings should not be considered interchangeable between the two systems.

Open-access reader

About this research paper

What this paper is about

Abstract Purpose To assess the repeatability and reproducibility of swept-source optical coherence tomography (SS-OCT) and Scheimpflug system and evaluate the agreement between the two systems in measuring multiple corneal regions in children. Methods Pachymetric and keratometric maps for both systems were evaluated. Central, midperipheral and peripheral corneal thickness (CT), keratometry and astigmatism power vectors were recorded. The three outcomes yielded by the same observer were used to assess intraobserver repeatability. The differences in the mean values provided by each observer were used to evaluate interobserver reproducibility. Within-subject standard deviation, test-retest repeatability (TRT) and coefficient of variation (CoV) were used to analyze the intraobserver repeatability and interobserver reproducibility. Paired T-test and Bland-Altman were used to appraise interdevice agreement. Results Seventy-eight eyes of 78 children were included. The CoV was ≤2.12 and 1.10%, respectively, for repeatability and reproducibility. TRT and CoV were lower for central and paracentral CT measurements than for peripheral measurements. The SS-OCT device generated higher precision when acquiring CT data, whereas Scheimpflug system showed higher reliability when measuring corneal keratometry. Although the CT readings measured using SS-OCT were significantly thinner than Scheimpflug device (P < 0.001), the central and thinnest CT values were still of high agreement. The interdevice agreement of keratometry measurement was high for the central corneal region and moderate for the paracentral and peripheral areas. Conclusions The precision of CT measurements by SS-OCT was higher, while the reliability of keratometry measurements by the Scheimpflug system was higher in children. Apart from the measured values in the central corneal region, the thickness and keratometry readings should not be considered interchangeable between the two systems.

Why it matters

OpenAlex reports 22 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Abstract Purpose To assess the repeatability and reproducibility of swept-source optical coherence tomography (SS-OCT) and Scheimpflug system and evaluate the agreement between the two systems in measuring multiple corneal regions in children. Methods Pachymetric and keratometric maps for both systems were evaluated. Central, midperipheral and peripheral corneal thickness (CT), keratometry and astigmatism power vectors were recorded. The three outcomes yielded by the same observer were used to assess intraobserver repeatability. The differences in the mean values provided by each observer were used to evaluate interobserver reproducibility. Within-subject standard deviation, test-retest repeatability (TRT) and coefficient of variation (CoV) were used to analyze the intraobserver repeatability and interobserver reproducibility. Paired T-test and Bland-Altman were used to appraise interdevice agreement. Results Seventy-eight eyes of 78 children were included. The CoV was ≤2.12 and 1.10%, respectively, for repeatability and reproducibility. TRT and CoV were lower for central and paracentral CT measurements than for peripheral measurements. The SS-OCT device generated higher precision when acquiring CT data, whereas Scheimpflug system showed higher reliability when measuring corneal keratometry. Although the CT readings measured using SS-OCT were significantly thinner than Scheimpflug device (P < 0.001), the central and thinnest CT values were still of high agreement. The interdevice agreement of keratometry measurement was high for the central corneal region and moderate for the paracentral and peripheral areas. Conclusions The precision of CT measurements by SS-OCT was higher, while the reliability of keratometry measurements by the Scheimpflug system was higher in children. Apart from the measured values in the central corneal region, the thickness and keratometry readings should not be considered interchangeable between the two systems.

Key concepts: Scheimpflug principle, Repeatability, Keratometer, Reproducibility, Intraclass correlation, Medicine, Ophthalmology, Coefficient of variation

Related papers

Back to paper searchBrowse research topicsOriginal source
The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging — Research Paper | ScholarLens