Low-dose CT coronaryangiography with prospective electrocardiogram-triggering using dual-source CT scanner
徐磊, 晏子旭, 张兆琪, Zhanming Fan, 吕飙, 于薇
Abstract
徐磊, 晏子旭, 张兆琪, Zhanming Fan, 吕飙, 于薇
Abstract
:Objective To evaluatelow-dose CT coronary angiography with prospective electrocardiogram (ECG)-triggering usingdual-source CT scanner.Methods Sixty-eight patients who underwent coronary CT angiographyusing a dual-source CT scanner were divided into 2 groups: group A (38 cases) and group B(30 cases).Prospective ECG-triggering sequence scan mode was employed for groupA.Inclusion criteria included: heart rate <70 bpm, sinus rhythm, and heart ratefluctuation less than 10 bpm.Data acquisition was set at 70% of theRR-interval.Retrospective ECG-gating helical scan was performed for group B.Inclusioncriteria included heart rates < 70 bpm and sinus rhythm.The exclusion criteria includedheart failure and serious arrhythmias.In both groups, patients with a BMI≥24 kg/m2 were examined with atube voltage of 120 kV, whereas patients with a BMI <24 kg/m2 were examined with a tubevoltage of 100 kV.All images were transferred to a workstation for further processing andanalysis.The imaging quality was evaluated.The imaging quality of coronary artery segmentswere compared with rank sum test between the two groups, and the radiation dose werecompared with t test.Results A total of 476 coronary artery segments were evaluated ingroup A and 372 segments were evaluated in group B.The mean score of imaging quality forcoronary artery segments in group A was 3.48±0.59 and that in group B was 3.53±0.58.Therewas no statistical difference in imaging quality between the two groups (Z=-1.432,P=0.187).The effective dose was on average (2.51±0.54) mSv (range 1.3--3.3 mSv) in groupA, whereas on average (14.55±3.54) rosy (range 7.1--20.2 mSv) in group B.There was astatistical difference between the two groups (t=18.484, P=0.000).Conclusions Low-doseprospective ECG-triggering sequence scan in dual-source CT coronary angiography isfeasible in patients with low heart rate and regular cardiac rhythm.This scan mode cansubstantially reduce radiation doses while preserving good diagnostic image quality. Key words: Tomography, X-ray computed; Coronary vessels; Radiationmonitoring
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
:Objective To evaluatelow-dose CT coronary angiography with prospective electrocardiogram (ECG)-triggering usingdual-source CT scanner.Methods Sixty-eight patients who underwent coronary CT angiographyusing a dual-source CT scanner were divided into 2 groups: group A (38 cases) and group B(30 cases).Prospective ECG-triggering sequence scan mode was employed for groupA.Inclusion criteria included: heart rate <70 bpm, sinus rhythm, and heart ratefluctuation less than 10 bpm.Data acquisition was set at 70% of theRR-interval.Retrospective ECG-gating helical scan was performed for group B.Inclusioncriteria included heart rates < 70 bpm and sinus rhythm.The exclusion criteria includedheart failure and serious arrhythmias.In both groups, patients with a BMI≥24 kg/m2 were examined with atube voltage of 120 kV, whereas patients with a BMI <24 kg/m2 were examined with a tubevoltage of 100 kV.All images were transferred to a workstation for further processing andanalysis.The imaging quality was evaluated.The imaging quality of coronary artery segmentswere compared with rank sum test between the two groups, and the radiation dose werecompared with t test.Results A total of 476 coronary artery segments were evaluated ingroup A and 372 segments were evaluated in group B.The mean score of imaging quality forcoronary artery segments in group A was 3.48±0.59 and that in group B was 3.53±0.58.Therewas no statistical difference in imaging quality between the two groups (Z=-1.432,P=0.187).The effective dose was on average (2.51±0.54) mSv (range 1.3--3.3 mSv) in groupA, whereas on average (14.55±3.54) rosy (range 7.1--20.2 mSv) in group B.There was astatistical difference between the two groups (t=18.484, P=0.000).Conclusions Low-doseprospective ECG-triggering sequence scan in dual-source CT coronary angiography isfeasible in patients with low heart rate and regular cardiac rhythm.This scan mode cansubstantially reduce radiation doses while preserving good diagnostic image quality. Key words: Tomography, X-ray computed; Coronary vessels; Radiationmonitoring
Key concepts: Medicine, Prospective cohort study, Sinus rhythm, Nuclear medicine, Helical scan, Artery, Image quality, Heart rate