A study on radiation dose and image quality of chest spectral imaging and conventional CT in anthropomorphic phantom

Qiuzhen Xu, Hailei Shao, Yan Lyu

Open publisher page 0 citations

Abstract

Objective To obtain the parameters of optimal spectral scanning mode in chest spectral CT imaging, radiation doses, image quality, CNR and subjective scores are compared in spectral and conventional scan modes with anthropomorphic chest phantom. Methods An anthropomorphic chest phantom underwent both conventional scan and spectral scan using three different protocols: GSI assist mode (protocol A); GSI mode with mean tube current (protocol B); GSI mode with maximum tube current(protocol C). All above scans were performed with the noise index (NI) as 9 and 11 respectively. The radiation dose was recorded, and the SD values of adipose tissue and muscle tissue of 5 different sections were measured in regions of interest (ROIs) to evaluate the image quality. All scans were subjectively scored in 5 score system by two senior radiologists on lung markings and branches with the lung window technique. Results The effective dose (E)values of conventional CT scan and protocols A, B and C were 8.0, 8.5, 6.2, 10.4 and 5.3, 5.1, 4.3, 6.2 mSv. When NI=9, the difference of SD values had statistically significance between conventional scan and protocols A, C (F=4.496, P 0.05). The difference of CNR and subjective score had no statistically significance in either the same NI different scanning modes or the same scanning mode with different NIs. Conclusions Appropriate spectral scanning mode had no significant difference with conventional CT scanning in radiation dose, yet higher image quality can be obtained. Furthermore, appropriate NI can reduce the radiation dose obviously while the similar image quality was achieved. Considering the two factors, GSI assist mode can achieve balance between radiation dose and image quality. Key words: Anthropomorphic chest phantom; Spectral CT; Conventional CT; Image quality; Radiation dose

About this research paper

What this paper is about

Objective To obtain the parameters of optimal spectral scanning mode in chest spectral CT imaging, radiation doses, image quality, CNR and subjective scores are compared in spectral and conventional scan modes with anthropomorphic chest phantom. Methods An anthropomorphic chest phantom underwent both conventional scan and spectral scan using three different protocols: GSI assist mode (protocol A); GSI mode with mean tube current (protocol B); GSI mode with maximum tube current(protocol C). All above scans were performed with the noise index (NI) as 9 and 11 respectively. The radiation dose was recorded, and the SD values of adipose tissue and muscle tissue of 5 different sections were measured in regions of interest (ROIs) to evaluate the image quality. All scans were subjectively scored in 5 score system by two senior radiologists on lung markings and branches with the lung window technique. Results The effective dose (E)values of conventional CT scan and protocols A, B and C were 8.0, 8.5, 6.2, 10.4 and 5.3, 5.1, 4.3, 6.2 mSv. When NI=9, the difference of SD values had statistically significance between conventional scan and protocols A, C (F=4.496, P 0.05). The difference of CNR and subjective score had no statistically significance in either the same NI different scanning modes or the same scanning mode with different NIs. Conclusions Appropriate spectral scanning mode had no significant difference with conventional CT scanning in radiation dose, yet higher image quality can be obtained. Furthermore, appropriate NI can reduce the radiation dose obviously while the similar image quality was achieved. Considering the two factors, GSI assist mode can achieve balance between radiation dose and image quality. Key words: Anthropomorphic chest phantom; Spectral CT; Conventional CT; Image quality; Radiation dose

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To obtain the parameters of optimal spectral scanning mode in chest spectral CT imaging, radiation doses, image quality, CNR and subjective scores are compared in spectral and conventional scan modes with anthropomorphic chest phantom. Methods An anthropomorphic chest phantom underwent both conventional scan and spectral scan using three different protocols: GSI assist mode (protocol A); GSI mode with mean tube current (protocol B); GSI mode with maximum tube current(protocol C). All above scans were performed with the noise index (NI) as 9 and 11 respectively. The radiation dose was recorded, and the SD values of adipose tissue and muscle tissue of 5 different sections were measured in regions of interest (ROIs) to evaluate the image quality. All scans were subjectively scored in 5 score system by two senior radiologists on lung markings and branches with the lung window technique. Results The effective dose (E)values of conventional CT scan and protocols A, B and C were 8.0, 8.5, 6.2, 10.4 and 5.3, 5.1, 4.3, 6.2 mSv. When NI=9, the difference of SD values had statistically significance between conventional scan and protocols A, C (F=4.496, P 0.05). The difference of CNR and subjective score had no statistically significance in either the same NI different scanning modes or the same scanning mode with different NIs. Conclusions Appropriate spectral scanning mode had no significant difference with conventional CT scanning in radiation dose, yet higher image quality can be obtained. Furthermore, appropriate NI can reduce the radiation dose obviously while the similar image quality was achieved. Considering the two factors, GSI assist mode can achieve balance between radiation dose and image quality. Key words: Anthropomorphic chest phantom; Spectral CT; Conventional CT; Image quality; Radiation dose

Key concepts: Imaging phantom, Image quality, Nuclear medicine, Spectral imaging, Image noise, Radiation dose, Medicine, Helical scan