A comparison of 20-MHz and 50-MHz ultrasonography in imaging and measuring the skin

Menglong Ran, Dehua Liu, Jingqiu Zhang, Ping Tu, Shu-xia Yang

Open publisher page 0 citations

Abstract

Objective To investigate imaging characteristics of human skin in different sites by using 20-MHz and 50-MHz high-frequency ultrasonography, and to compare the reliability of skin thickness measurement by the above two approaches of ultrasonography. Methods A total of 39 healthy volunteers aged 18-39 years were enrolled into this study. Then, 20-MHz and 50-MHz ultrasonography were separately performed to image 20 different sites on the body, and the thickness of the epidermis and dermis were measured and evaluated by 3 dermatologists independently. The ultrasonic images were analyzed, and the intraclass correlation coefficient (ICC) was used to assess and compare the reliability of skin thickness measurement by 20-MHz and 50-MHz ultrasound. Results The ICC values were less than or close to 0.7 in epidermal thickness measurement at almost all the tested sites between 20-MHz and 50-MHz ultrasono-graphy, suggesting poor reliability. However, 20-MHz and 50-MHz ultrasonography showed excellent reliability in dermal thickness measurement with the ICC greater than 0.75 at almost all the tested smooth and flat body sites. The 20-MHz ultrasound could provide clear images of the dermis and subcutaneous tissue, while 50-MHz ultrasound could only provide clearer images of the dermis in details. In some uneven sites with complex anatomical structures, 20-MHz and 50-MHz ultrasound probes had their own advantages and disadvantages in imaging depth and detail resolution. Conclusions The 20-MHz and 50-MHz ultrasono-graphy both can serve as non-invasive imaging techniques to show structures of the epidermis and dermis better. In different parts of human body, ultrasound frequencies should be selected according to their imaging characteristics. Key words: Ultrasonography; Skin structure; Skin thickness; Interclass correlation coefficient

About this research paper

What this paper is about

Objective To investigate imaging characteristics of human skin in different sites by using 20-MHz and 50-MHz high-frequency ultrasonography, and to compare the reliability of skin thickness measurement by the above two approaches of ultrasonography. Methods A total of 39 healthy volunteers aged 18-39 years were enrolled into this study. Then, 20-MHz and 50-MHz ultrasonography were separately performed to image 20 different sites on the body, and the thickness of the epidermis and dermis were measured and evaluated by 3 dermatologists independently. The ultrasonic images were analyzed, and the intraclass correlation coefficient (ICC) was used to assess and compare the reliability of skin thickness measurement by 20-MHz and 50-MHz ultrasound. Results The ICC values were less than or close to 0.7 in epidermal thickness measurement at almost all the tested sites between 20-MHz and 50-MHz ultrasono-graphy, suggesting poor reliability. However, 20-MHz and 50-MHz ultrasonography showed excellent reliability in dermal thickness measurement with the ICC greater than 0.75 at almost all the tested smooth and flat body sites. The 20-MHz ultrasound could provide clear images of the dermis and subcutaneous tissue, while 50-MHz ultrasound could only provide clearer images of the dermis in details. In some uneven sites with complex anatomical structures, 20-MHz and 50-MHz ultrasound probes had their own advantages and disadvantages in imaging depth and detail resolution. Conclusions The 20-MHz and 50-MHz ultrasono-graphy both can serve as non-invasive imaging techniques to show structures of the epidermis and dermis better. In different parts of human body, ultrasound frequencies should be selected according to their imaging characteristics. Key words: Ultrasonography; Skin structure; Skin thickness; Interclass correlation coefficient

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To investigate imaging characteristics of human skin in different sites by using 20-MHz and 50-MHz high-frequency ultrasonography, and to compare the reliability of skin thickness measurement by the above two approaches of ultrasonography. Methods A total of 39 healthy volunteers aged 18-39 years were enrolled into this study. Then, 20-MHz and 50-MHz ultrasonography were separately performed to image 20 different sites on the body, and the thickness of the epidermis and dermis were measured and evaluated by 3 dermatologists independently. The ultrasonic images were analyzed, and the intraclass correlation coefficient (ICC) was used to assess and compare the reliability of skin thickness measurement by 20-MHz and 50-MHz ultrasound. Results The ICC values were less than or close to 0.7 in epidermal thickness measurement at almost all the tested sites between 20-MHz and 50-MHz ultrasono-graphy, suggesting poor reliability. However, 20-MHz and 50-MHz ultrasonography showed excellent reliability in dermal thickness measurement with the ICC greater than 0.75 at almost all the tested smooth and flat body sites. The 20-MHz ultrasound could provide clear images of the dermis and subcutaneous tissue, while 50-MHz ultrasound could only provide clearer images of the dermis in details. In some uneven sites with complex anatomical structures, 20-MHz and 50-MHz ultrasound probes had their own advantages and disadvantages in imaging depth and detail resolution. Conclusions The 20-MHz and 50-MHz ultrasono-graphy both can serve as non-invasive imaging techniques to show structures of the epidermis and dermis better. In different parts of human body, ultrasound frequencies should be selected according to their imaging characteristics. Key words: Ultrasonography; Skin structure; Skin thickness; Interclass correlation coefficient

Key concepts: Ultrasound, Dermis, Intraclass correlation, Ultrasonography, Biomedical engineering, Skin thickness, Reliability (semiconductor), Ultrasonic sensor