Consistency in measuring corneal curvature with Sirius corneal topography and a keratometer

Su-rong Luo, Wei Chen

Open publisher page 0 citations

Abstract

Objective To explore the differences and agreements between Sirius corneal topography and a keratometer in the measurement of corneal curvature before laser in situ keratomileusis (LASIK) in patients with myopia.Methods This was a prospective control study.One hundred and seventy eyes of 85 patients with myopia were selected at The People's Hospital of Shaoxing.Measurements of corneal curvature were taken with both Sirius corneal topography and a keratometer.A paired samples t-test and Pearson linear correlation were used for statistical analysis,and a BlandAltman analysis was used to evaluate the consistency of the two methods of measurement.Results The corneal curvature K1 values measured by Sirius corneal topography and a keratometer were 42.41±1.25 D and 42.33±1.28 D,K2 values were 43.36±1.38 D and 43.56±1.38 D,and (K1+K2)/2 values were 42.88±1.28 D and 42.92±1.29 D,respectively,indicating significant differences (t1=5.241,P<0.01; t2=-9.936,P<0.01; taverage=-3.021,P<0.01) and significant linear correlations between K1 and K2 (r1=0.989,P<0.01 ;r2=0.981,P<0.01; raverage=0.990,P<0.01).The differences between Sirius corneal topography and keratometry measurements of corneal curvature for K1 and K2 values and the differences in average corneal curvature (K1+K2)/2 were 0.08 D,-0.20 D and-0.04 D,respectively.The differences between standard deviations were 0.1897 D,0.2644 D and 0.1790 D and the 95% limits of agreement were (0.45 D,-0.30 D),(0.32 D,-0.72 D) and (0.31 D,-0.39 D).There were 2.4%(4/170),3.5(6/170) and 4.7%(8/170) points for the 95% limits of agreement,and consistency within

About this research paper

What this paper is about

Objective To explore the differences and agreements between Sirius corneal topography and a keratometer in the measurement of corneal curvature before laser in situ keratomileusis (LASIK) in patients with myopia.Methods This was a prospective control study.One hundred and seventy eyes of 85 patients with myopia were selected at The People's Hospital of Shaoxing.Measurements of corneal curvature were taken with both Sirius corneal topography and a keratometer.A paired samples t-test and Pearson linear correlation were used for statistical analysis,and a BlandAltman analysis was used to evaluate the consistency of the two methods of measurement.Results The corneal curvature K1 values measured by Sirius corneal topography and a keratometer were 42.41±1.25 D and 42.33±1.28 D,K2 values were 43.36±1.38 D and 43.56±1.38 D,and (K1+K2)/2 values were 42.88±1.28 D and 42.92±1.29 D,respectively,indicating significant differences (t1=5.241,P<0.01; t2=-9.936,P<0.01; taverage=-3.021,P<0.01) and significant linear correlations between K1 and K2 (r1=0.989,P<0.01 ;r2=0.981,P<0.01; raverage=0.990,P<0.01).The differences between Sirius corneal topography and keratometry measurements of corneal curvature for K1 and K2 values and the differences in average corneal curvature (K1+K2)/2 were 0.08 D,-0.20 D and-0.04 D,respectively.The differences between standard deviations were 0.1897 D,0.2644 D and 0.1790 D and the 95% limits of agreement were (0.45 D,-0.30 D),(0.32 D,-0.72 D) and (0.31 D,-0.39 D).There were 2.4%(4/170),3.5(6/170) and 4.7%(8/170) points for the 95% limits of agreement,and consistency within

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To explore the differences and agreements between Sirius corneal topography and a keratometer in the measurement of corneal curvature before laser in situ keratomileusis (LASIK) in patients with myopia.Methods This was a prospective control study.One hundred and seventy eyes of 85 patients with myopia were selected at The People's Hospital of Shaoxing.Measurements of corneal curvature were taken with both Sirius corneal topography and a keratometer.A paired samples t-test and Pearson linear correlation were used for statistical analysis,and a BlandAltman analysis was used to evaluate the consistency of the two methods of measurement.Results The corneal curvature K1 values measured by Sirius corneal topography and a keratometer were 42.41±1.25 D and 42.33±1.28 D,K2 values were 43.36±1.38 D and 43.56±1.38 D,and (K1+K2)/2 values were 42.88±1.28 D and 42.92±1.29 D,respectively,indicating significant differences (t1=5.241,P<0.01; t2=-9.936,P<0.01; taverage=-3.021,P<0.01) and significant linear correlations between K1 and K2 (r1=0.989,P<0.01 ;r2=0.981,P<0.01; raverage=0.990,P<0.01).The differences between Sirius corneal topography and keratometry measurements of corneal curvature for K1 and K2 values and the differences in average corneal curvature (K1+K2)/2 were 0.08 D,-0.20 D and-0.04 D,respectively.The differences between standard deviations were 0.1897 D,0.2644 D and 0.1790 D and the 95% limits of agreement were (0.45 D,-0.30 D),(0.32 D,-0.72 D) and (0.31 D,-0.39 D).There were 2.4%(4/170),3.5(6/170) and 4.7%(8/170) points for the 95% limits of agreement,and consistency within

Key concepts: Keratometer, Sirius, Keratomileusis, Ophthalmology, LASIK, Corneal topography, Medicine, Cornea

Related papers

Back to paper searchBrowse research topicsOriginal source
Consistency in measuring corneal curvature with Sirius corneal topography and a keratometer — Research Paper | ScholarLens