Overranging and overbeaming measurement in area detector computed tomography: A method for simultaneous measurement in volume helical acquisition

Atsushi Urikura, T. Hara, Tsukasa Yoshida, Eiji Nishimaru, Takashi Hoshino, Yoshihiro Nakaya, Masahiro Endo

Open full text 19 citations

Abstract

PURPOSE: We propose a novel method to assess overbeaming and overranging, as well as the effect of reducing longitudinal exposure range, by using a dynamic z-collimator in area detector computed tomography. METHODS AND MATERIALS: A 500-mm diameter cylindrical imaging plate was exposed by helical scanning in a dark room. The beam collimation of the helical acquisitions was set at 32 and 80 mm. Overbeaming and overranging with the dynamic z-collimator were measured. RESULTS: The actual beam widths were approximately 39 and 88 mm at 32 and 80 mm collimation, respectively, and were relatively reduced owing to increased beam collimation. Overranging was 27.0 and 48.2 mm with a pitch of 0.83 and 1.49 at 32 mm collimation and 72.5 and 83.1 mm with a pitch of 0.87 and 0.99 at 80 mm collimation. The dynamic z-collimator relatively reduced the overranging by 17.3% and 17.1% for the 32 and 80 mm collimation, respectively. CONCLUSION: We devised a method to simultaneously measure overbeaming and overranging with only one helical acquisition. Although the dynamic z-collimator reduced the overranging by approximately 17%, wider collimation widths and higher pitch settings would increase the exposure dose outside the scan range.

Open-access reader

About this research paper

What this paper is about

PURPOSE: We propose a novel method to assess overbeaming and overranging, as well as the effect of reducing longitudinal exposure range, by using a dynamic z-collimator in area detector computed tomography. METHODS AND MATERIALS: A 500-mm diameter cylindrical imaging plate was exposed by helical scanning in a dark room. The beam collimation of the helical acquisitions was set at 32 and 80 mm. Overbeaming and overranging with the dynamic z-collimator were measured. RESULTS: The actual beam widths were approximately 39 and 88 mm at 32 and 80 mm collimation, respectively, and were relatively reduced owing to increased beam collimation. Overranging was 27.0 and 48.2 mm with a pitch of 0.83 and 1.49 at 32 mm collimation and 72.5 and 83.1 mm with a pitch of 0.87 and 0.99 at 80 mm collimation. The dynamic z-collimator relatively reduced the overranging by 17.3% and 17.1% for the 32 and 80 mm collimation, respectively. CONCLUSION: We devised a method to simultaneously measure overbeaming and overranging with only one helical acquisition. Although the dynamic z-collimator reduced the overranging by approximately 17%, wider collimation widths and higher pitch settings would increase the exposure dose outside the scan range.

Why it matters

OpenAlex reports 19 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

PURPOSE: We propose a novel method to assess overbeaming and overranging, as well as the effect of reducing longitudinal exposure range, by using a dynamic z-collimator in area detector computed tomography. METHODS AND MATERIALS: A 500-mm diameter cylindrical imaging plate was exposed by helical scanning in a dark room. The beam collimation of the helical acquisitions was set at 32 and 80 mm. Overbeaming and overranging with the dynamic z-collimator were measured. RESULTS: The actual beam widths were approximately 39 and 88 mm at 32 and 80 mm collimation, respectively, and were relatively reduced owing to increased beam collimation. Overranging was 27.0 and 48.2 mm with a pitch of 0.83 and 1.49 at 32 mm collimation and 72.5 and 83.1 mm with a pitch of 0.87 and 0.99 at 80 mm collimation. The dynamic z-collimator relatively reduced the overranging by 17.3% and 17.1% for the 32 and 80 mm collimation, respectively. CONCLUSION: We devised a method to simultaneously measure overbeaming and overranging with only one helical acquisition. Although the dynamic z-collimator reduced the overranging by approximately 17%, wider collimation widths and higher pitch settings would increase the exposure dose outside the scan range.

Key concepts: Collimator, Collimated light, Optics, Nuclear medicine, Detector, Beam (structure), Materials science, Physics

Related papers

Back to paper searchBrowse research topicsOriginal source
Overranging and overbeaming measurement in area detector computed tomography: A method for simultaneous measurement in volume helical acquisition — Research Paper | ScholarLens