Validated method for trace impurities analysis in bulk gas using Gas Chromatography with Pulse Discharge Helium Ionization Detection

Ayu Hindayani, Oman Zuas, Muhammad Rizky Mulyana, Harry Budiman, Dyah Styarini, Fransiska Sri Herwahyu Krismastuti, Bernadus H. Sirenden

Open full text 3 citations

Abstract

Pulse discharge helium ionization detector (PDHID) is developed to respond the need of many applications in the area of chromatographic quantitative analysis. The PDHID is a general type of detector with very high detectability property covering a wide range of chemical compounds. Based on ISO/IEC 17025, a reliable and accurate of measurement result can be obtained by applying a validated method. This paper reports a validation study with regard to the application of GC-PDHID for trace impurity analysis in bulk gases. An optimized operating condition of GC was used to evaluate the method performance with parameters including precision (repeatability and reproducibility), linearity, limit of detection (LoD), and limit of quantification (LoQ). In this study, six trace compounds (CO 2 , H 2 , Ar, O 2 , CH 4 , and CO) in bulk gas were analyzed. The results show that the validated of the GC-PDHID method was repeatable and reproducible for the analysis of trace impurities in bulk gas. The %RSD of repeatability and reproducibility were found to be < 3.00% and < 10.50%, respectively. Good linearity of the validated method was obtained with the correlation coefficient (R 2 ) higher than 0.99 for all six trace compounds. In addition, the calculated LOD and LOQ of the GC-PDHID were found to be < 0.60 and < 2.00 μmol mol −1 respectively. In conclusion, validated method of the GC-PDHID for the measurement of the six trace compounds in bulk gas is reliable and fit for its purpose, implying that the validated method can be used in daily measurement in the testing laboratory.

Open-access reader

About this research paper

What this paper is about

Pulse discharge helium ionization detector (PDHID) is developed to respond the need of many applications in the area of chromatographic quantitative analysis. The PDHID is a general type of detector with very high detectability property covering a wide range of chemical compounds. Based on ISO/IEC 17025, a reliable and accurate of measurement result can be obtained by applying a validated method. This paper reports a validation study with regard to the application of GC-PDHID for trace impurity analysis in bulk gases. An optimized operating condition of GC was used to evaluate the method performance with parameters including precision (repeatability and reproducibility), linearity, limit of detection (LoD), and limit of quantification (LoQ). In this study, six trace compounds (CO 2 , H 2 , Ar, O 2 , CH 4 , and CO) in bulk gas were analyzed. The results show that the validated of the GC-PDHID method was repeatable and reproducible for the analysis of trace impurities in bulk gas. The %RSD of repeatability and reproducibility were found to be < 3.00% and < 10.50%, respectively. Good linearity of the validated method was obtained with the correlation coefficient (R 2 ) higher than 0.99 for all six trace compounds. In addition, the calculated LOD and LOQ of the GC-PDHID were found to be < 0.60 and < 2.00 μmol mol −1 respectively. In conclusion, validated method of the GC-PDHID for the measurement of the six trace compounds in bulk gas is reliable and fit for its purpose, implying that the validated method can be used in daily measurement in the testing laboratory.

Why it matters

OpenAlex reports 3 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Pulse discharge helium ionization detector (PDHID) is developed to respond the need of many applications in the area of chromatographic quantitative analysis. The PDHID is a general type of detector with very high detectability property covering a wide range of chemical compounds. Based on ISO/IEC 17025, a reliable and accurate of measurement result can be obtained by applying a validated method. This paper reports a validation study with regard to the application of GC-PDHID for trace impurity analysis in bulk gases. An optimized operating condition of GC was used to evaluate the method performance with parameters including precision (repeatability and reproducibility), linearity, limit of detection (LoD), and limit of quantification (LoQ). In this study, six trace compounds (CO 2 , H 2 , Ar, O 2 , CH 4 , and CO) in bulk gas were analyzed. The results show that the validated of the GC-PDHID method was repeatable and reproducible for the analysis of trace impurities in bulk gas. The %RSD of repeatability and reproducibility were found to be < 3.00% and < 10.50%, respectively. Good linearity of the validated method was obtained with the correlation coefficient (R 2 ) higher than 0.99 for all six trace compounds. In addition, the calculated LOD and LOQ of the GC-PDHID were found to be < 0.60 and < 2.00 μmol mol −1 respectively. In conclusion, validated method of the GC-PDHID for the measurement of the six trace compounds in bulk gas is reliable and fit for its purpose, implying that the validated method can be used in daily measurement in the testing laboratory.

Key concepts: Repeatability, Reproducibility, Detection limit, Analytical Chemistry (journal), Gas chromatography, Helium, Flame ionization detector, Impurity

Related papers

Back to paper searchBrowse research topicsOriginal source
Validated method for trace impurities analysis in bulk gas using Gas Chromatography with Pulse Discharge Helium Ionization Detection — Research Paper | ScholarLens