The feasibility of fingerstick blood collection for point-of-care HIV-1 viral load monitoring in rural Zambia

Catherine Gayle Sutcliffe, Kara M. Palamountain, Sylvia Maunga, Kelly M. Searle, Philip E. Thuma, William John Moss, Mark J. Fisher

Open full text 8 citations

Abstract

Viral load monitoring for HIV treatment is recommended but not feasible in many settings. A point-of-care test using capillary blood would increase access but may require up to 200 μL of blood to achieve a lower limit of detection of 1000 copies/mL. This cross-sectional study evaluated the feasibility of collecting 200 μL of capillary blood and blood collection preferences among adults in rural Zambia. Adults seeking HIV counseling and testing at Macha Hospital were recruited in 2015. Capillary blood was collected in four 50 μL tubes. Blood collection was categorized as complete (200 μL collected), partial (all tubes filled but <200 μL obtained due to collection techniques), or incomplete (1-4 tubes attempted; <200 μL obtained due to insufficient blood flow). One fingerstick was required for 90% of the 201 participants. A median blood volume of 196 μL was collected. Complete, partial and incomplete collection was achieved in 34%, 59% and 6% of participants. The majority of participants (95%) preferred fingerstick over venous blood collection. A point-of-care viral load test requiring up to 200 μL of blood is feasible in a rural setting but would require training and supervision to ensure that sufficient blood was collected.

About this research paper

What this paper is about

Viral load monitoring for HIV treatment is recommended but not feasible in many settings. A point-of-care test using capillary blood would increase access but may require up to 200 μL of blood to achieve a lower limit of detection of 1000 copies/mL. This cross-sectional study evaluated the feasibility of collecting 200 μL of capillary blood and blood collection preferences among adults in rural Zambia. Adults seeking HIV counseling and testing at Macha Hospital were recruited in 2015. Capillary blood was collected in four 50 μL tubes. Blood collection was categorized as complete (200 μL collected), partial (all tubes filled but <200 μL obtained due to collection techniques), or incomplete (1-4 tubes attempted; <200 μL obtained due to insufficient blood flow). One fingerstick was required for 90% of the 201 participants. A median blood volume of 196 μL was collected. Complete, partial and incomplete collection was achieved in 34%, 59% and 6% of participants. The majority of participants (95%) preferred fingerstick over venous blood collection. A point-of-care viral load test requiring up to 200 μL of blood is feasible in a rural setting but would require training and supervision to ensure that sufficient blood was collected.

Why it matters

OpenAlex reports 8 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Viral load monitoring for HIV treatment is recommended but not feasible in many settings. A point-of-care test using capillary blood would increase access but may require up to 200 μL of blood to achieve a lower limit of detection of 1000 copies/mL. This cross-sectional study evaluated the feasibility of collecting 200 μL of capillary blood and blood collection preferences among adults in rural Zambia. Adults seeking HIV counseling and testing at Macha Hospital were recruited in 2015. Capillary blood was collected in four 50 μL tubes. Blood collection was categorized as complete (200 μL collected), partial (all tubes filled but <200 μL obtained due to collection techniques), or incomplete (1-4 tubes attempted; <200 μL obtained due to insufficient blood flow). One fingerstick was required for 90% of the 201 participants. A median blood volume of 196 μL was collected. Complete, partial and incomplete collection was achieved in 34%, 59% and 6% of participants. The majority of participants (95%) preferred fingerstick over venous blood collection. A point-of-care viral load test requiring up to 200 μL of blood is feasible in a rural setting but would require training and supervision to ensure that sufficient blood was collected.