2018Journal of Interventional Cardiac ElectrophysiologyOpen access

The learning curve associated with the implantation of the Nanostim leadless pacemaker

on behalf of the Leadless II IDE and, Fleur V.Y. Tjong, Observational Study Investigators, Niek E. G. Beurskens, Petr Neužil, Pascal Defaye, Peter-Paul Delnoy, John Ip, Juan J. García Guerrero, Mayer Rashtian, Rajesh Banker, Vivek Y. Reddy, Derek V. Exner, Johannes Sperzel, Reinoud E. Knops

Open full text 19 citations

Abstract

PURPOSE: Use of novel medical technologies, such as leadless pacemaker (LP) therapy, may be subjected to a learning curve effect. The objective of the current study was to assess the impact of operators' experience on the occurrence of serious adverse device effects (SADE) and procedural efficiency. METHODS: Patients implanted with a Nanostim LP (Abbott, USA) within two prospective studies (i.e., LEADLESS ll IDE and Leadless Observational Study) were assessed. Patients were categorized into quartiles based on operator experience. Learning curve analysis included the comparison of SADE rates at 30 days post-implant per quartile and between patients in quartile 4 (> 10 implants) and patients in quartiles 1 through 3 (1-10 implants). Procedural efficiency was assessed based on procedure duration and repositioning attempts. RESULTS: Nanostim LP implant was performed in 1439 patients by 171 implanters at 60 centers in 10 countries. A total of 91 (6.4%) patients experienced a SADE in the first 30 days. SADE rates dropped from 7.4 to 4.5% (p = 0.038) after more than 10 implants per operator. Total procedure duration decreased from 30.9 ± 19.1 min in quartile 1 to 21.6 ± 13.2 min (p < 0.001) in quartile 4. The need for multiple repositionings during the LP procedure reduced in quartile 4 (14.8%), compared to quartiles 1 (26.8%; p < 0.001), 2 (26.6%; p < 0.001), and 3 (20.4%; p = 0.03). CONCLUSIONS: Learning curves exist for Nanostim LP implantation. Procedure efficiency improved with increased operator experience, according to a decrease in the incidence of SADE, procedure duration, and repositioning attempts.

About this research paper

What this paper is about

PURPOSE: Use of novel medical technologies, such as leadless pacemaker (LP) therapy, may be subjected to a learning curve effect. The objective of the current study was to assess the impact of operators' experience on the occurrence of serious adverse device effects (SADE) and procedural efficiency. METHODS: Patients implanted with a Nanostim LP (Abbott, USA) within two prospective studies (i.e., LEADLESS ll IDE and Leadless Observational Study) were assessed. Patients were categorized into quartiles based on operator experience. Learning curve analysis included the comparison of SADE rates at 30 days post-implant per quartile and between patients in quartile 4 (> 10 implants) and patients in quartiles 1 through 3 (1-10 implants). Procedural efficiency was assessed based on procedure duration and repositioning attempts. RESULTS: Nanostim LP implant was performed in 1439 patients by 171 implanters at 60 centers in 10 countries. A total of 91 (6.4%) patients experienced a SADE in the first 30 days. SADE rates dropped from 7.4 to 4.5% (p = 0.038) after more than 10 implants per operator. Total procedure duration decreased from 30.9 ± 19.1 min in quartile 1 to 21.6 ± 13.2 min (p < 0.001) in quartile 4. The need for multiple repositionings during the LP procedure reduced in quartile 4 (14.8%), compared to quartiles 1 (26.8%; p < 0.001), 2 (26.6%; p < 0.001), and 3 (20.4%; p = 0.03). CONCLUSIONS: Learning curves exist for Nanostim LP implantation. Procedure efficiency improved with increased operator experience, according to a decrease in the incidence of SADE, procedure duration, and repositioning attempts.

Why it matters

OpenAlex reports 19 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

PURPOSE: Use of novel medical technologies, such as leadless pacemaker (LP) therapy, may be subjected to a learning curve effect. The objective of the current study was to assess the impact of operators' experience on the occurrence of serious adverse device effects (SADE) and procedural efficiency. METHODS: Patients implanted with a Nanostim LP (Abbott, USA) within two prospective studies (i.e., LEADLESS ll IDE and Leadless Observational Study) were assessed. Patients were categorized into quartiles based on operator experience. Learning curve analysis included the comparison of SADE rates at 30 days post-implant per quartile and between patients in quartile 4 (> 10 implants) and patients in quartiles 1 through 3 (1-10 implants). Procedural efficiency was assessed based on procedure duration and repositioning attempts. RESULTS: Nanostim LP implant was performed in 1439 patients by 171 implanters at 60 centers in 10 countries. A total of 91 (6.4%) patients experienced a SADE in the first 30 days. SADE rates dropped from 7.4 to 4.5% (p = 0.038) after more than 10 implants per operator. Total procedure duration decreased from 30.9 ± 19.1 min in quartile 1 to 21.6 ± 13.2 min (p < 0.001) in quartile 4. The need for multiple repositionings during the LP procedure reduced in quartile 4 (14.8%), compared to quartiles 1 (26.8%; p < 0.001), 2 (26.6%; p < 0.001), and 3 (20.4%; p = 0.03). CONCLUSIONS: Learning curves exist for Nanostim LP implantation. Procedure efficiency improved with increased operator experience, according to a decrease in the incidence of SADE, procedure duration, and repositioning attempts.

Key concepts: Medicine, Learning curve, Cardiology, Internal medicine, Economics, Management

Related papers

Back to paper searchBrowse research topicsOriginal source
The learning curve associated with the implantation of the Nanostim leadless pacemaker — Research Paper | ScholarLens