Independent evaluation of the effectiveness of IsoCal in improving image center accuracy on Varian TrueBeam and Clinac machines

Weiliang Du, Song Gao, Wei Jiang, Rajat J. Kudchadker

Open full text 14 citations

Abstract

Modern medical linear accelerators (linacs) are often equipped with image guidance systems that are capable of megavolt (MV), kilovolt (kV), planar, or volumetric imaging. On Varian TrueBeam linacs, the isocenter accuracies of the imaging systems are calibrated with a procedure named IsoCal. On Clinac series linacs from Varian, installation of IsoCal is optional and the effects of IsoCal on the imaging systems can be turned on or off after the IsoCal procedure is performed. In this study, we report on the effectiveness of IsoCal in improving the coincidence of the image centers with the radiation isocenter, using an independent Winston-Lutz (WL) method to locate the radiation isocenter. A ball-bearing phantom was imaged with 2D MV, 2D kV, and cone beam computed radiography systems on two TrueBeam and two Clinac machines. Using the same phantom, digital WL tests with 16 combinations of gantry and collimator angles were performed to locate the radiation isocenter. The offsets between the IsoCal-calibrated image centers and the WL radiation isocenter were found to be within 0.4 mm on the four linacs in this study. When IsoCal was turned off, the maximal offsets of the image centers were greater than 1.0 mm on the two Clinac machines. The method developed in this study can be used as a vendor-independent quality assurance tool to assess the isocentricity of the image centers and radiation central axes.

Open-access reader

About this research paper

What this paper is about

Modern medical linear accelerators (linacs) are often equipped with image guidance systems that are capable of megavolt (MV), kilovolt (kV), planar, or volumetric imaging. On Varian TrueBeam linacs, the isocenter accuracies of the imaging systems are calibrated with a procedure named IsoCal. On Clinac series linacs from Varian, installation of IsoCal is optional and the effects of IsoCal on the imaging systems can be turned on or off after the IsoCal procedure is performed. In this study, we report on the effectiveness of IsoCal in improving the coincidence of the image centers with the radiation isocenter, using an independent Winston-Lutz (WL) method to locate the radiation isocenter. A ball-bearing phantom was imaged with 2D MV, 2D kV, and cone beam computed radiography systems on two TrueBeam and two Clinac machines. Using the same phantom, digital WL tests with 16 combinations of gantry and collimator angles were performed to locate the radiation isocenter. The offsets between the IsoCal-calibrated image centers and the WL radiation isocenter were found to be within 0.4 mm on the four linacs in this study. When IsoCal was turned off, the maximal offsets of the image centers were greater than 1.0 mm on the two Clinac machines. The method developed in this study can be used as a vendor-independent quality assurance tool to assess the isocentricity of the image centers and radiation central axes.

Why it matters

OpenAlex reports 14 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Modern medical linear accelerators (linacs) are often equipped with image guidance systems that are capable of megavolt (MV), kilovolt (kV), planar, or volumetric imaging. On Varian TrueBeam linacs, the isocenter accuracies of the imaging systems are calibrated with a procedure named IsoCal. On Clinac series linacs from Varian, installation of IsoCal is optional and the effects of IsoCal on the imaging systems can be turned on or off after the IsoCal procedure is performed. In this study, we report on the effectiveness of IsoCal in improving the coincidence of the image centers with the radiation isocenter, using an independent Winston-Lutz (WL) method to locate the radiation isocenter. A ball-bearing phantom was imaged with 2D MV, 2D kV, and cone beam computed radiography systems on two TrueBeam and two Clinac machines. Using the same phantom, digital WL tests with 16 combinations of gantry and collimator angles were performed to locate the radiation isocenter. The offsets between the IsoCal-calibrated image centers and the WL radiation isocenter were found to be within 0.4 mm on the four linacs in this study. When IsoCal was turned off, the maximal offsets of the image centers were greater than 1.0 mm on the two Clinac machines. The method developed in this study can be used as a vendor-independent quality assurance tool to assess the isocentricity of the image centers and radiation central axes.

Key concepts: Truebeam, Isocenter, Linear particle accelerator, Imaging phantom, Quality assurance, Collimator, Nuclear medicine, Dosimetry

Related papers

Back to paper searchBrowse research topicsOriginal source
Independent evaluation of the effectiveness of IsoCal in improving image center accuracy on Varian TrueBeam and Clinac machines — Research Paper | ScholarLens