Evaluation of a new photon-counting imaging detector (PCD) with various acquisition modes
Alok Shankar, Jordan M. Krebs, Daniel R. Bednarek, Stephen Rudin
Abstract
Alok Shankar, Jordan M. Krebs, Daniel R. Bednarek, Stephen Rudin
Abstract
The prospect of improved low noise, high speed, and dual-energy imaging that may be associated with the use of photon-counting imaging detectors (PCD) has motivated this evaluation of a newly upgraded version of a prototype PCD. The XCounter Actaeon was evaluated in its four acquisition modes each based upon varying signal processing firmware including a mode with charge sharing correction that enables neighboring pixels that share the energy from one incident x-ray photon detection to be counted only once at the proper summed energy in the pixel with the largest charge deposition. Since this PCD is a CdTe-based direct detector with 100 μm pixels, such charge sharing for typical medical x-ray energy photons may occur frequently and must be corrected to achieve more accurate counts. This charge sharing correction is achieved with an Anti-Coincidence Circuit (ACC) which prevents double pixel counting from one event as well as prevents counting from either event if they are below a preset threshold. Various physical parameters of the PCD were evaluated including linearity, sensitivity, pulse pile-up effects, dark noise, spatial resolution, noise power spectrum, and detective quantum efficiency.
OpenAlex reports 9 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The prospect of improved low noise, high speed, and dual-energy imaging that may be associated with the use of photon-counting imaging detectors (PCD) has motivated this evaluation of a newly upgraded version of a prototype PCD. The XCounter Actaeon was evaluated in its four acquisition modes each based upon varying signal processing firmware including a mode with charge sharing correction that enables neighboring pixels that share the energy from one incident x-ray photon detection to be counted only once at the proper summed energy in the pixel with the largest charge deposition. Since this PCD is a CdTe-based direct detector with 100 μm pixels, such charge sharing for typical medical x-ray energy photons may occur frequently and must be corrected to achieve more accurate counts. This charge sharing correction is achieved with an Anti-Coincidence Circuit (ACC) which prevents double pixel counting from one event as well as prevents counting from either event if they are below a preset threshold. Various physical parameters of the PCD were evaluated including linearity, sensitivity, pulse pile-up effects, dark noise, spatial resolution, noise power spectrum, and detective quantum efficiency.
Key concepts: Photon counting, Detector, Photon, Physics, Computer science, Optics