Investigation of Exposure Factors for Various Breast Composition and Thicknesses in Digital Screening Mammography Related to Breast Dose

Khaled Alkhalifah, Ajit Brindhaban

Open full text 6 citations

Abstract

OBJECTIVE: To investigate the effect of exposure factors used in digital screening mammography on image quality of different breast compositions. MATERIAL AND METHODS: A digital mammography unit, with tungsten (W) as target, rhodium (Rh) and silver (Ag) as filters, and amorphous selenium detectors, was used to image Computerized Imaging Reference Systems (CIRS) Model 12A phantoms of thickness 4, 5, and 6 cm. Images of each phantom were obtained using target-filter combinations of W/Rh and W/Ag, at 28, 30, and 32 kVp. Images were evaluated by 5 senior technologists with experience in mammography. Image scores were assigned, for each type of feature present in the phantom. Statistical analysis was performed using nonparametric tests to compare sets of image scores at p = 0.05. RESULTS: A small but statistically significant improvement was detected in the visibility of microcalcifications (8.8 ± 0.2; p = 0.031) for the W/Rh combination but this did not show any differences in the visibility of masses or fibers. The entrance skin dose (ESD) and mean glandular dose (MGD) were lower for the W/Ag (ESD = 1.30-3.70; MGD = 0.44-0.93 mGy) combination compared to W/Rh (ESD = 1.66-5.40; MGD = 0.52-1.12 mGy). The Mann-Whitney test revealed that 30-kV exposure with the W/Rh combination showed a significantly better visibility of specks in the 30/70 phantom compared to other exposures. CONCLUSION: The use of an Rh filter showed a better image quality for all phantoms. 28 and 30 kVp with the W/Rh combination provided a slightly better image quality, and the MGD is less than 1.2 mGy.

About this research paper

What this paper is about

OBJECTIVE: To investigate the effect of exposure factors used in digital screening mammography on image quality of different breast compositions. MATERIAL AND METHODS: A digital mammography unit, with tungsten (W) as target, rhodium (Rh) and silver (Ag) as filters, and amorphous selenium detectors, was used to image Computerized Imaging Reference Systems (CIRS) Model 12A phantoms of thickness 4, 5, and 6 cm. Images of each phantom were obtained using target-filter combinations of W/Rh and W/Ag, at 28, 30, and 32 kVp. Images were evaluated by 5 senior technologists with experience in mammography. Image scores were assigned, for each type of feature present in the phantom. Statistical analysis was performed using nonparametric tests to compare sets of image scores at p = 0.05. RESULTS: A small but statistically significant improvement was detected in the visibility of microcalcifications (8.8 ± 0.2; p = 0.031) for the W/Rh combination but this did not show any differences in the visibility of masses or fibers. The entrance skin dose (ESD) and mean glandular dose (MGD) were lower for the W/Ag (ESD = 1.30-3.70; MGD = 0.44-0.93 mGy) combination compared to W/Rh (ESD = 1.66-5.40; MGD = 0.52-1.12 mGy). The Mann-Whitney test revealed that 30-kV exposure with the W/Rh combination showed a significantly better visibility of specks in the 30/70 phantom compared to other exposures. CONCLUSION: The use of an Rh filter showed a better image quality for all phantoms. 28 and 30 kVp with the W/Rh combination provided a slightly better image quality, and the MGD is less than 1.2 mGy.

Why it matters

OpenAlex reports 6 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

OBJECTIVE: To investigate the effect of exposure factors used in digital screening mammography on image quality of different breast compositions. MATERIAL AND METHODS: A digital mammography unit, with tungsten (W) as target, rhodium (Rh) and silver (Ag) as filters, and amorphous selenium detectors, was used to image Computerized Imaging Reference Systems (CIRS) Model 12A phantoms of thickness 4, 5, and 6 cm. Images of each phantom were obtained using target-filter combinations of W/Rh and W/Ag, at 28, 30, and 32 kVp. Images were evaluated by 5 senior technologists with experience in mammography. Image scores were assigned, for each type of feature present in the phantom. Statistical analysis was performed using nonparametric tests to compare sets of image scores at p = 0.05. RESULTS: A small but statistically significant improvement was detected in the visibility of microcalcifications (8.8 ± 0.2; p = 0.031) for the W/Rh combination but this did not show any differences in the visibility of masses or fibers. The entrance skin dose (ESD) and mean glandular dose (MGD) were lower for the W/Ag (ESD = 1.30-3.70; MGD = 0.44-0.93 mGy) combination compared to W/Rh (ESD = 1.66-5.40; MGD = 0.52-1.12 mGy). The Mann-Whitney test revealed that 30-kV exposure with the W/Rh combination showed a significantly better visibility of specks in the 30/70 phantom compared to other exposures. CONCLUSION: The use of an Rh filter showed a better image quality for all phantoms. 28 and 30 kVp with the W/Rh combination provided a slightly better image quality, and the MGD is less than 1.2 mGy.

Key concepts: Medicine, Mammography, Digital mammography, Imaging phantom, Nuclear medicine, Image quality, Breast cancer, Internal medicine