Mercury versus Oscillometric Device in Measuring Blood Pressure in Elderly: Which Is Authentic?

V. Sujanitha, T. Kumanan‬, Mahesan Guruparan, Thavakularatnam Inthuja

Open full text 0 citations

Abstract

Aim: To compare the systolic and diastolic blood pressure in both arms in elderly patients by mercury sphygmomanometer and oscillometric recorders adopting standard practice of measuring blood pressure.Materials and methods: It is a cross sectional study done in 204 elderly patients aged more than 60 years over a period of two months at teaching hospital, Jaffna, Sri Lanka.Blood pressure was measured in both arms by both devices adopting the standard practice of blood pressure measurements.Results: The mean Systolic Blood Pressure (SBP) by oscillometric recorder on both right and left arm was 4.18mmHg and 3.66 mmHg higher than mercury sphygmomanometer readings.The mean Diastolic Blood Pressure (DBP) by oscillometric recorder on both right arm and left arm was 1.67 and 1.64 mmHg lower than mercury sphygmomanometer readings.There wasn't any significant difference in inter arm blood pressure measurement.The study group consisted of 54.3% males and 45.7% of females.The mean age of the study group was 69.46 years.Conclusion: Though there is no much difference observed in the gross values in reading the blood pressure by both oscillometric recorders and mercury sphygmomanometers, there is statistically significant discrepancy in blood pressure readings particularly in the elderly population which needs to be studied further in detail.

Open-access reader

About this research paper

What this paper is about

Aim: To compare the systolic and diastolic blood pressure in both arms in elderly patients by mercury sphygmomanometer and oscillometric recorders adopting standard practice of measuring blood pressure.Materials and methods: It is a cross sectional study done in 204 elderly patients aged more than 60 years over a period of two months at teaching hospital, Jaffna, Sri Lanka.Blood pressure was measured in both arms by both devices adopting the standard practice of blood pressure measurements.Results: The mean Systolic Blood Pressure (SBP) by oscillometric recorder on both right and left arm was 4.18mmHg and 3.66 mmHg higher than mercury sphygmomanometer readings.The mean Diastolic Blood Pressure (DBP) by oscillometric recorder on both right arm and left arm was 1.67 and 1.64 mmHg lower than mercury sphygmomanometer readings.There wasn't any significant difference in inter arm blood pressure measurement.The study group consisted of 54.3% males and 45.7% of females.The mean age of the study group was 69.46 years.Conclusion: Though there is no much difference observed in the gross values in reading the blood pressure by both oscillometric recorders and mercury sphygmomanometers, there is statistically significant discrepancy in blood pressure readings particularly in the elderly population which needs to be studied further in detail.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Aim: To compare the systolic and diastolic blood pressure in both arms in elderly patients by mercury sphygmomanometer and oscillometric recorders adopting standard practice of measuring blood pressure.Materials and methods: It is a cross sectional study done in 204 elderly patients aged more than 60 years over a period of two months at teaching hospital, Jaffna, Sri Lanka.Blood pressure was measured in both arms by both devices adopting the standard practice of blood pressure measurements.Results: The mean Systolic Blood Pressure (SBP) by oscillometric recorder on both right and left arm was 4.18mmHg and 3.66 mmHg higher than mercury sphygmomanometer readings.The mean Diastolic Blood Pressure (DBP) by oscillometric recorder on both right arm and left arm was 1.67 and 1.64 mmHg lower than mercury sphygmomanometer readings.There wasn't any significant difference in inter arm blood pressure measurement.The study group consisted of 54.3% males and 45.7% of females.The mean age of the study group was 69.46 years.Conclusion: Though there is no much difference observed in the gross values in reading the blood pressure by both oscillometric recorders and mercury sphygmomanometers, there is statistically significant discrepancy in blood pressure readings particularly in the elderly population which needs to be studied further in detail.

Key concepts: Mercury (programming language), Blood pressure, Environmental science, Medicine, Internal medicine, Computer science, Programming language