Stability assessment of radiation isocenter with the gimbaled linac system

Hideharu Miura, Shuichi Ozawa, Shintaro Tsuda, Kiyoshi Yamada, Yasushi Nagata

Open full text 1 citations

Abstract

Purpose: We report the results of our year-long radiation isocenter accuracy verification for daily quality assurance (QA) implementation on a Vero4DRT system. Methods: The radiation isocenter was calculated using a cube phantom with a steel ball of diameter 10 mm fixed to the center of the phantom. A single photon beam was set with a field size of 100 × 100 mm 2 . Coincidence of the centroid of the steel ball at kiloVolt X-ray imaging isocenter and megaVolt beam radiation isocenter at each gantry and ring angle was tested. This procedure was performed for gantry angles of 0°, 90°, 180°, and 270°, and ring angles of 0°, 20°, and 340°. The centroid of the steel ball and the center of the radiation field were calculated to analyze the radiation isocenter error. This analysis was automatically calculated using the Daily Check tool in the Vero4DRT system. This QA was implemented between 24 August 2015 and 23 August 2016. Results: The average and standard deviation for pan and tilt directions were 0.12 ± 0.10 mm and -0.20 ± 0.13 mm, respectively. The maximum radiation isocenter accuracy error was 0.50 mm in both directions. Conclusion: The radiation isocenter alignment for the one year duration of the experiment was performed with high accuracy.

Open-access reader

About this research paper

What this paper is about

Purpose: We report the results of our year-long radiation isocenter accuracy verification for daily quality assurance (QA) implementation on a Vero4DRT system. Methods: The radiation isocenter was calculated using a cube phantom with a steel ball of diameter 10 mm fixed to the center of the phantom. A single photon beam was set with a field size of 100 × 100 mm 2 . Coincidence of the centroid of the steel ball at kiloVolt X-ray imaging isocenter and megaVolt beam radiation isocenter at each gantry and ring angle was tested. This procedure was performed for gantry angles of 0°, 90°, 180°, and 270°, and ring angles of 0°, 20°, and 340°. The centroid of the steel ball and the center of the radiation field were calculated to analyze the radiation isocenter error. This analysis was automatically calculated using the Daily Check tool in the Vero4DRT system. This QA was implemented between 24 August 2015 and 23 August 2016. Results: The average and standard deviation for pan and tilt directions were 0.12 ± 0.10 mm and -0.20 ± 0.13 mm, respectively. The maximum radiation isocenter accuracy error was 0.50 mm in both directions. Conclusion: The radiation isocenter alignment for the one year duration of the experiment was performed with high accuracy.

Why it matters

OpenAlex reports 1 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Purpose: We report the results of our year-long radiation isocenter accuracy verification for daily quality assurance (QA) implementation on a Vero4DRT system. Methods: The radiation isocenter was calculated using a cube phantom with a steel ball of diameter 10 mm fixed to the center of the phantom. A single photon beam was set with a field size of 100 × 100 mm 2 . Coincidence of the centroid of the steel ball at kiloVolt X-ray imaging isocenter and megaVolt beam radiation isocenter at each gantry and ring angle was tested. This procedure was performed for gantry angles of 0°, 90°, 180°, and 270°, and ring angles of 0°, 20°, and 340°. The centroid of the steel ball and the center of the radiation field were calculated to analyze the radiation isocenter error. This analysis was automatically calculated using the Daily Check tool in the Vero4DRT system. This QA was implemented between 24 August 2015 and 23 August 2016. Results: The average and standard deviation for pan and tilt directions were 0.12 ± 0.10 mm and -0.20 ± 0.13 mm, respectively. The maximum radiation isocenter accuracy error was 0.50 mm in both directions. Conclusion: The radiation isocenter alignment for the one year duration of the experiment was performed with high accuracy.

Key concepts: Isocenter, Imaging phantom, Linear particle accelerator, Optics, Nuclear medicine, Physics, Radiation, Centroid