Enhanced imaging of metallic knee prostheses by using a representative phantom
R.P.A. Boekestijn, R.J. van Dijk, Tim M.N. van Helden, R. Hoogland
Abstract
R.P.A. Boekestijn, R.J. van Dijk, Tim M.N. van Helden, R. Hoogland
Abstract
Objective: Investigate how a representative phantom can contribute to enhanced imaging of a Total Knee Arthroplasty (TKA) prosthesis in low-field MRI. Method: Ingredients of the phantom are gelatine (gelling agent), Dotarem® (T1 modifier), agarose (T2 modifier) and demineralised water. The T1 and T2 relaxation times of tissue mimicking samples are measured using a spin echo sequence in a 0.25 Tesla low field MRI. Artefacts sizes in the MR images of the phantom materials with prosthesis are measured and compared with the artefact sizes in the MR images of a human knee with prosthesis. Results: Relaxation times of the samples ranged from 210 to 480 for T1. It was not possible to calculate T2 values in the samples. The artefact sizes differ from artefacts in a real knee with prosthesis. Relaxation times of the human knee ranged between 160 and 470 for T1 and between 25 and 127 for T2. The phantom is not representative for a human knee. Comparison between phantom with prosthesis and knee with prosthesis show conflicting results. Especially the FSE X-MAR scan of the phantom shows more artefacts than the standard FSE scan. Conclusion: It was not possible to create a representative phantom. It is seen that Dotarem® and agarose change T1 relaxation times. The metal artefacts do not show consistent results. A representative phantom with complex structures should determine in what way metal artefacts can be reduced, to better image TKA problems.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective: Investigate how a representative phantom can contribute to enhanced imaging of a Total Knee Arthroplasty (TKA) prosthesis in low-field MRI. Method: Ingredients of the phantom are gelatine (gelling agent), Dotarem® (T1 modifier), agarose (T2 modifier) and demineralised water. The T1 and T2 relaxation times of tissue mimicking samples are measured using a spin echo sequence in a 0.25 Tesla low field MRI. Artefacts sizes in the MR images of the phantom materials with prosthesis are measured and compared with the artefact sizes in the MR images of a human knee with prosthesis. Results: Relaxation times of the samples ranged from 210 to 480 for T1. It was not possible to calculate T2 values in the samples. The artefact sizes differ from artefacts in a real knee with prosthesis. Relaxation times of the human knee ranged between 160 and 470 for T1 and between 25 and 127 for T2. The phantom is not representative for a human knee. Comparison between phantom with prosthesis and knee with prosthesis show conflicting results. Especially the FSE X-MAR scan of the phantom shows more artefacts than the standard FSE scan. Conclusion: It was not possible to create a representative phantom. It is seen that Dotarem® and agarose change T1 relaxation times. The metal artefacts do not show consistent results. A representative phantom with complex structures should determine in what way metal artefacts can be reduced, to better image TKA problems.
Key concepts: Imaging phantom, Prosthesis, T2 relaxation, Biomedical engineering, Materials science, Knee prosthesis, Agarose, Total knee arthroplasty