[Quantitative blood flow measurements with the c. w.-Doppler ultrasound procedure. An animal experiment study].

Hildebrandt Vj

Open publisher page 0 citations

Abstract

The qualification of the c.w.-ultrasound-doppler method for quantitative blood flow measurement was investigated in an animal experimental study in dogs. The simultaneous measurements by means of the electromagnetic flow measurement showed a correlation coefficient from 0.865 to 0.942 in dependence on the exposure angle to ultrasonic waves. The c.w.-ultrasound-doppler method appears suitable for quantitative blood flow measurements after these results. The exactness of the definition of the perpendicular as reference angle for deriving the exposure angle for ultrasonic waves has proved true.

About this research paper

What this paper is about

The qualification of the c.w.-ultrasound-doppler method for quantitative blood flow measurement was investigated in an animal experimental study in dogs. The simultaneous measurements by means of the electromagnetic flow measurement showed a correlation coefficient from 0.865 to 0.942 in dependence on the exposure angle to ultrasonic waves. The c.w.-ultrasound-doppler method appears suitable for quantitative blood flow measurements after these results. The exactness of the definition of the perpendicular as reference angle for deriving the exposure angle for ultrasonic waves has proved true.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

The qualification of the c.w.-ultrasound-doppler method for quantitative blood flow measurement was investigated in an animal experimental study in dogs. The simultaneous measurements by means of the electromagnetic flow measurement showed a correlation coefficient from 0.865 to 0.942 in dependence on the exposure angle to ultrasonic waves. The c.w.-ultrasound-doppler method appears suitable for quantitative blood flow measurements after these results. The exactness of the definition of the perpendicular as reference angle for deriving the exposure angle for ultrasonic waves has proved true.

Key concepts: Doppler effect, Blood flow, Ultrasound, Ultrasonic sensor, Doppler ultrasound, Perpendicular, Acoustics, Ultrasonic flow meter

Related papers

Back to paper searchBrowse research topicsOriginal source
[Quantitative blood flow measurements with the c. w.-Doppler ultrasound procedure. An animal experiment study]. — Research Paper | ScholarLens