[Analyze of the residue level of acrylic acid in cosmetics by ion chromatography method].
Haiying Yu, Qiyan Li, Xiaobing Wang, Junjie Li, Wenchao Gao
Abstract
Haiying Yu, Qiyan Li, Xiaobing Wang, Junjie Li, Wenchao Gao
Abstract
OBJECTIVE: To develop a method for acrylic acid analysis in cosmetics by ion chromatography. METHODS: Cosmetic samples were extracted by ultrasonic wave, then the separation was carried out on IonPac AS11-HC column (4 mm x 250 mm I. D.) with 30 mmol/L KOH as the eluent. The flow rate of the eluent was 0.8 ml/min. ASRS 300 4 - mm anion inhibitor and conductivity detector were used. Standard curve method was used for quantitative analysis. RESULTS: The linear ranges of calibration curve was 1.16 - 46.4 mg/L for acrylic acid, r = 0. 9996. The limit of detection was 4.8 mg/kg. The recoveries of acrylic acid with three different base materials were 89.1% - 92.6%, 93.5% - 96.2% and 90.5% - 94.8%. The intra-day and the inter-day precision were 1.8% and 5.8%, respectively. CONCLUSION: The established ion chromatographic method was simple, speedy, sensitive, and can be used to analyze the residue level of acrylic acid in cosmetics.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
OBJECTIVE: To develop a method for acrylic acid analysis in cosmetics by ion chromatography. METHODS: Cosmetic samples were extracted by ultrasonic wave, then the separation was carried out on IonPac AS11-HC column (4 mm x 250 mm I. D.) with 30 mmol/L KOH as the eluent. The flow rate of the eluent was 0.8 ml/min. ASRS 300 4 - mm anion inhibitor and conductivity detector were used. Standard curve method was used for quantitative analysis. RESULTS: The linear ranges of calibration curve was 1.16 - 46.4 mg/L for acrylic acid, r = 0. 9996. The limit of detection was 4.8 mg/kg. The recoveries of acrylic acid with three different base materials were 89.1% - 92.6%, 93.5% - 96.2% and 90.5% - 94.8%. The intra-day and the inter-day precision were 1.8% and 5.8%, respectively. CONCLUSION: The established ion chromatographic method was simple, speedy, sensitive, and can be used to analyze the residue level of acrylic acid in cosmetics.
Key concepts: Calibration curve, Detection limit, Chromatography, Column chromatography, Ion chromatography, Acrylic acid, Residue (chemistry), Chemistry