[Difference in measurements of erythrocyte sedimentation rate obtained using two devices of the same model in a patient with marked leukocytosis].

Takahiro Nojiri, Akiko Masuda, Ryunosuke Ohkawa, Kiyoko Koido, Hironori Shimosaka, Shigeo Okubo, Hiromitsu Yokota, Makoto Kurano, Yasuhito Nannya, Hitoshi Ikeda, Mineo Kurokawa, Yutaka Yatomi

Open publisher page 0 citations

Abstract

The erythrocyte sedimentation rate (ESR) has been used as an index for inflammatory conditions, such as infectious diseases, autoimmune diseases, and malignancies. The ESR values of a 37-year-old male with marked leukocytosis due to chronic myeloid leukemia showed remarkable differences between two devices of the same model (Ves-Matic 30, DIESSE Diagnostica Senese). From the appearance of the tested tube after the ESR measurement, the values obtained using one device might have been falsely low, whereas the values obtained using the other device were likely to have been accurate. The difference of the ESR values between the two devices might have occurred by the false detection of transmitted light during the transition from the erythrocyte layer to the leukocyte layer. These findings suggest that in cases with marked leukocytosis the accuracy of ESR should be confirmed with the appearance of the test tube.

About this research paper

What this paper is about

The erythrocyte sedimentation rate (ESR) has been used as an index for inflammatory conditions, such as infectious diseases, autoimmune diseases, and malignancies. The ESR values of a 37-year-old male with marked leukocytosis due to chronic myeloid leukemia showed remarkable differences between two devices of the same model (Ves-Matic 30, DIESSE Diagnostica Senese). From the appearance of the tested tube after the ESR measurement, the values obtained using one device might have been falsely low, whereas the values obtained using the other device were likely to have been accurate. The difference of the ESR values between the two devices might have occurred by the false detection of transmitted light during the transition from the erythrocyte layer to the leukocyte layer. These findings suggest that in cases with marked leukocytosis the accuracy of ESR should be confirmed with the appearance of the test tube.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

The erythrocyte sedimentation rate (ESR) has been used as an index for inflammatory conditions, such as infectious diseases, autoimmune diseases, and malignancies. The ESR values of a 37-year-old male with marked leukocytosis due to chronic myeloid leukemia showed remarkable differences between two devices of the same model (Ves-Matic 30, DIESSE Diagnostica Senese). From the appearance of the tested tube after the ESR measurement, the values obtained using one device might have been falsely low, whereas the values obtained using the other device were likely to have been accurate. The difference of the ESR values between the two devices might have occurred by the false detection of transmitted light during the transition from the erythrocyte layer to the leukocyte layer. These findings suggest that in cases with marked leukocytosis the accuracy of ESR should be confirmed with the appearance of the test tube.

Key concepts: Leukocytosis, Erythrocyte sedimentation rate, Myeloid leukemia, Medicine, Myeloid, Sedimentation, Pathology, Significant difference

Related papers

Back to paper searchBrowse research topicsOriginal source
[Difference in measurements of erythrocyte sedimentation rate obtained using two devices of the same model in a patient with marked leukocytosis]. — Research Paper | ScholarLens