Blood Cholesterol Concentration Measured by CR3000: Fingerstick versus Venous Sampling

Valeriana Teresa Sblendorio, Beniamino Palmieri, Graziano Riccioni

Open publisher page 9 citations

Abstract

Currently, the clinical practice of desktop or Point of Care (PoC) analyzers for lipid measurements has gained wide popularity. Designed to quickly perform measurements on microlitre(microL) quantities of blood, these instruments can be used in non-laboratory settings, such as physicians offices or field-testing sites and can provide measurements in whole blood, serum, or plasma, using either venous or capillary blood samples. The aim of this study is to examine the relationship between cholesterol determinations in venous and capillary samples using the CR3000 PoC system. The study was performed on 21 unselected adult volunteers, and no exclusion criteria was adopted. The mean cholesterol concentration for the venous blood samples measured was 164 mg/dL. The values obtained in the capillary blood samples averaged 168 mg/dL, which is only slightly higher (e.g., 2.87%) than the venous sample measurements. Moreover, the total variance was statistically similar for venous and capillary measurements (F value = 1.199, where the upper critical value of the F distribution is 2.124, p < 0.05). The results of our study support the concept that CR3000 total cholesterol testing can be performed safely and accurately in either venous or capillary specimens.

About this research paper

What this paper is about

Currently, the clinical practice of desktop or Point of Care (PoC) analyzers for lipid measurements has gained wide popularity. Designed to quickly perform measurements on microlitre(microL) quantities of blood, these instruments can be used in non-laboratory settings, such as physicians offices or field-testing sites and can provide measurements in whole blood, serum, or plasma, using either venous or capillary blood samples. The aim of this study is to examine the relationship between cholesterol determinations in venous and capillary samples using the CR3000 PoC system. The study was performed on 21 unselected adult volunteers, and no exclusion criteria was adopted. The mean cholesterol concentration for the venous blood samples measured was 164 mg/dL. The values obtained in the capillary blood samples averaged 168 mg/dL, which is only slightly higher (e.g., 2.87%) than the venous sample measurements. Moreover, the total variance was statistically similar for venous and capillary measurements (F value = 1.199, where the upper critical value of the F distribution is 2.124, p < 0.05). The results of our study support the concept that CR3000 total cholesterol testing can be performed safely and accurately in either venous or capillary specimens.

Why it matters

OpenAlex reports 9 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Currently, the clinical practice of desktop or Point of Care (PoC) analyzers for lipid measurements has gained wide popularity. Designed to quickly perform measurements on microlitre(microL) quantities of blood, these instruments can be used in non-laboratory settings, such as physicians offices or field-testing sites and can provide measurements in whole blood, serum, or plasma, using either venous or capillary blood samples. The aim of this study is to examine the relationship between cholesterol determinations in venous and capillary samples using the CR3000 PoC system. The study was performed on 21 unselected adult volunteers, and no exclusion criteria was adopted. The mean cholesterol concentration for the venous blood samples measured was 164 mg/dL. The values obtained in the capillary blood samples averaged 168 mg/dL, which is only slightly higher (e.g., 2.87%) than the venous sample measurements. Moreover, the total variance was statistically similar for venous and capillary measurements (F value = 1.199, where the upper critical value of the F distribution is 2.124, p < 0.05). The results of our study support the concept that CR3000 total cholesterol testing can be performed safely and accurately in either venous or capillary specimens.

Key concepts: Fingerstick, Venous blood, Medicine, Point of care, Whole blood, Internal medicine, Pathology, Endocrinology

Related papers

Back to paper searchBrowse research topicsOriginal source
Blood Cholesterol Concentration Measured by CR3000: Fingerstick versus Venous Sampling — Research Paper | ScholarLens