[Scanning electron microscopy studies of morphologic changes in ultra-high molecular weight polyethylene materials after various processing steps].

Jörg Jerosch, Sabine Fuchs, Tannin A. Schmidt, R. Reichelt

Open publisher page 0 citations

Abstract

UNLABELLED: Scanning electron microscopic (SEM) studies were carried out on various orthopaedic implants made of ultra-high molecular weight polyethylene (UHMWPE). Both non-sterilized and gamma-sterilized machined (milled) implants were used. Milled implants showed significant surface irregularities, and cross-sections also revealed irregularities in deeper layers of the polyethylene. In gamma-sterilized implants, cracks measuring 10-30 microns in length were found at a depth of 1.5 to 2.5 mm. Crater-like defects were also occasionally seen. CLINICAL RELEVANCE: On the basis of these findings, direct compression moulding and a sterilization process in which oxygen is eliminated can be recommended as ways of improving the quality of UHMWPE used for orthopaedic implants.

About this research paper

What this paper is about

UNLABELLED: Scanning electron microscopic (SEM) studies were carried out on various orthopaedic implants made of ultra-high molecular weight polyethylene (UHMWPE). Both non-sterilized and gamma-sterilized machined (milled) implants were used. Milled implants showed significant surface irregularities, and cross-sections also revealed irregularities in deeper layers of the polyethylene. In gamma-sterilized implants, cracks measuring 10-30 microns in length were found at a depth of 1.5 to 2.5 mm. Crater-like defects were also occasionally seen. CLINICAL RELEVANCE: On the basis of these findings, direct compression moulding and a sterilization process in which oxygen is eliminated can be recommended as ways of improving the quality of UHMWPE used for orthopaedic implants.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

UNLABELLED: Scanning electron microscopic (SEM) studies were carried out on various orthopaedic implants made of ultra-high molecular weight polyethylene (UHMWPE). Both non-sterilized and gamma-sterilized machined (milled) implants were used. Milled implants showed significant surface irregularities, and cross-sections also revealed irregularities in deeper layers of the polyethylene. In gamma-sterilized implants, cracks measuring 10-30 microns in length were found at a depth of 1.5 to 2.5 mm. Crater-like defects were also occasionally seen. CLINICAL RELEVANCE: On the basis of these findings, direct compression moulding and a sterilization process in which oxygen is eliminated can be recommended as ways of improving the quality of UHMWPE used for orthopaedic implants.

Key concepts: Ultra-high-molecular-weight polyethylene, Polyethylene, Scanning electron microscope, Materials science, Composite material, Sterilization (economics), Biomedical engineering, Medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
[Scanning electron microscopy studies of morphologic changes in ultra-high molecular weight polyethylene materials after various processing steps]. — Research Paper | ScholarLens