[Scanning electron microscopy studies of morphologic changes in ultra-high molecular weight polyethylene materials after various processing steps].
Jörg Jerosch, Sabine Fuchs, Tannin A. Schmidt, R. Reichelt
Abstract
Jörg Jerosch, Sabine Fuchs, Tannin A. Schmidt, R. Reichelt
Abstract
UNLABELLED: Scanning electron microscopic (SEM) studies were carried out on various orthopaedic implants made of ultra-high molecular weight polyethylene (UHMWPE). Both non-sterilized and gamma-sterilized machined (milled) implants were used. Milled implants showed significant surface irregularities, and cross-sections also revealed irregularities in deeper layers of the polyethylene. In gamma-sterilized implants, cracks measuring 10-30 microns in length were found at a depth of 1.5 to 2.5 mm. Crater-like defects were also occasionally seen. CLINICAL RELEVANCE: On the basis of these findings, direct compression moulding and a sterilization process in which oxygen is eliminated can be recommended as ways of improving the quality of UHMWPE used for orthopaedic implants.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
UNLABELLED: Scanning electron microscopic (SEM) studies were carried out on various orthopaedic implants made of ultra-high molecular weight polyethylene (UHMWPE). Both non-sterilized and gamma-sterilized machined (milled) implants were used. Milled implants showed significant surface irregularities, and cross-sections also revealed irregularities in deeper layers of the polyethylene. In gamma-sterilized implants, cracks measuring 10-30 microns in length were found at a depth of 1.5 to 2.5 mm. Crater-like defects were also occasionally seen. CLINICAL RELEVANCE: On the basis of these findings, direct compression moulding and a sterilization process in which oxygen is eliminated can be recommended as ways of improving the quality of UHMWPE used for orthopaedic implants.
Key concepts: Ultra-high-molecular-weight polyethylene, Polyethylene, Scanning electron microscope, Materials science, Composite material, Sterilization (economics), Biomedical engineering, Medicine