[Reproducibility of measurements with the Heidelberg retina tomograph in fundus elevations].

P. Janknecht, Jens Funk, L. L. Hansen

Open publisher page 6 citations

Abstract

PURPOSE: The Heidelberg Retina Tomograph (HRT) was intended to evaluate the optic nerve head. It is, however, also possible to measure retinal elevations. There are no systemic data on the reproducibility of HRT measurements of retinal elevations. We now provide such data. METHODS: 117 measurements in patients with diseases which are characterized by retinal elevations were analyzed with the HRT. We measured the "volume above surface" of the retinal elevations. Each patient was analyzed three times, and the coefficient of variation was calculated. RESULTS: The coefficient of variation was between 0.55% and 40.5%. Some 42% of the coefficients of variation were between 0% and 10%, 31% between 5% and 10%, 21% between 10% and 20%, and 6% larger than 20%. CONCLUSION: The reproducibility of the HRT measurements in retinal elevations is clinically sufficient. The relation between sufficient reproducibility and accuracy is discussed.

About this research paper

What this paper is about

PURPOSE: The Heidelberg Retina Tomograph (HRT) was intended to evaluate the optic nerve head. It is, however, also possible to measure retinal elevations. There are no systemic data on the reproducibility of HRT measurements of retinal elevations. We now provide such data. METHODS: 117 measurements in patients with diseases which are characterized by retinal elevations were analyzed with the HRT. We measured the "volume above surface" of the retinal elevations. Each patient was analyzed three times, and the coefficient of variation was calculated. RESULTS: The coefficient of variation was between 0.55% and 40.5%. Some 42% of the coefficients of variation were between 0% and 10%, 31% between 5% and 10%, 21% between 10% and 20%, and 6% larger than 20%. CONCLUSION: The reproducibility of the HRT measurements in retinal elevations is clinically sufficient. The relation between sufficient reproducibility and accuracy is discussed.

Why it matters

OpenAlex reports 6 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

PURPOSE: The Heidelberg Retina Tomograph (HRT) was intended to evaluate the optic nerve head. It is, however, also possible to measure retinal elevations. There are no systemic data on the reproducibility of HRT measurements of retinal elevations. We now provide such data. METHODS: 117 measurements in patients with diseases which are characterized by retinal elevations were analyzed with the HRT. We measured the "volume above surface" of the retinal elevations. Each patient was analyzed three times, and the coefficient of variation was calculated. RESULTS: The coefficient of variation was between 0.55% and 40.5%. Some 42% of the coefficients of variation were between 0% and 10%, 31% between 5% and 10%, 21% between 10% and 20%, and 6% larger than 20%. CONCLUSION: The reproducibility of the HRT measurements in retinal elevations is clinically sufficient. The relation between sufficient reproducibility and accuracy is discussed.

Key concepts: Reproducibility, Coefficient of variation, Retinal, Ophthalmology, Fundus (uterus), Retina, Medicine, Optic nerve