2004•Japanese Journal of Radiological TechnologyOpen access

Optimization of the Chest Exposure Condition with a Contrast-detail Phantom : Evaluation of the Flat-panel Versus Computed Radiography Systems

Emi Kinoshita, Yoshiyuki Umezu, Kazuhisa Ogawa, Toyoyuki Katou, Hisao Arimura, Hideki Yoshikawa, Yoshiharu Higashida, Masafumi Ooki, Fukai Toyofuku

Open full text 3 citations

Abstract

In this study, we evaluated the performance of a digital chest imaging system using a contrast-detail (C-D) phantom. In the initial step, 76 sample images of the C-D phantom were produced by changing the doses from 0.5, 0.75, 1.0, 1.25, 1.5, to 2.0 times the dose for a screen-film (S/F) system. The sample images were analyzed by five radiological technologists and two medical physicists, and the image quality figure (IQF) was determined. The quality of each image was examined, and appropriate doses were determined from the calculated IQF to obtain the same image quality for other digital chest imaging systems. The method of determining IQF from C-D phantom analysis was very useful for comparing image quality and determining radiographic techniques.

Open-access reader

About this research paper

What this paper is about

In this study, we evaluated the performance of a digital chest imaging system using a contrast-detail (C-D) phantom. In the initial step, 76 sample images of the C-D phantom were produced by changing the doses from 0.5, 0.75, 1.0, 1.25, 1.5, to 2.0 times the dose for a screen-film (S/F) system. The sample images were analyzed by five radiological technologists and two medical physicists, and the image quality figure (IQF) was determined. The quality of each image was examined, and appropriate doses were determined from the calculated IQF to obtain the same image quality for other digital chest imaging systems. The method of determining IQF from C-D phantom analysis was very useful for comparing image quality and determining radiographic techniques.

Why it matters

OpenAlex reports 3 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

In this study, we evaluated the performance of a digital chest imaging system using a contrast-detail (C-D) phantom. In the initial step, 76 sample images of the C-D phantom were produced by changing the doses from 0.5, 0.75, 1.0, 1.25, 1.5, to 2.0 times the dose for a screen-film (S/F) system. The sample images were analyzed by five radiological technologists and two medical physicists, and the image quality figure (IQF) was determined. The quality of each image was examined, and appropriate doses were determined from the calculated IQF to obtain the same image quality for other digital chest imaging systems. The method of determining IQF from C-D phantom analysis was very useful for comparing image quality and determining radiographic techniques.

Key concepts: Imaging phantom, Radiography, Computed radiography, Contrast (vision), Nuclear medicine, Medicine, Flat panel, Radiology