[Dose distribution of photons and neutrons outside of the irradiation field of a 8-mev linear accelerator (author's transl)].
Nemec Hw, Jessica Roth
Abstract
Nemec Hw, Jessica Roth
Abstract
During the irradiation with a continuous radiation of 8 MeV by the linear accelerator LINAC SL 75/10, the dose distribution of photons and neutrons was measured outside of the irradiation field. The dose distribution of photons was mostly determined by thermoluminescence dose meters. There was a significant maximum of leak radiation on the back of the emitter head which showed a dose rate amounting to 7,5% of the dose rate of the central ray. Whereas the leak radiation on the patients' table was less than 0,1% of the dose of the central ray, the dose of scattered radiation emitted by the patient still amounts to 2% of the central ray dose in a distance of 15 cm from the field edge. The isodoses measured in the room demonstrate above all the importance of the lock for a sufficient radioprotection.--The dose rate of the neutrons formed in the shielding material by photonuclear reactions were measured mostly with uranium fission-track dosimeters. The highest equivalent dose rate was found on the back of the emitter head; it is about 17 rem/h. The maximum value of the equivalent dose rate of neutrons on the patient's table amounts to 1,4 rem/h and is almost independent of the size of the diaphragm. The isodoses in the room depend largely on the shape of the irradiation room; they confirm the existence of a sufficient radioprotection.
OpenAlex reports 2 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
During the irradiation with a continuous radiation of 8 MeV by the linear accelerator LINAC SL 75/10, the dose distribution of photons and neutrons was measured outside of the irradiation field. The dose distribution of photons was mostly determined by thermoluminescence dose meters. There was a significant maximum of leak radiation on the back of the emitter head which showed a dose rate amounting to 7,5% of the dose rate of the central ray. Whereas the leak radiation on the patients' table was less than 0,1% of the dose of the central ray, the dose of scattered radiation emitted by the patient still amounts to 2% of the central ray dose in a distance of 15 cm from the field edge. The isodoses measured in the room demonstrate above all the importance of the lock for a sufficient radioprotection.--The dose rate of the neutrons formed in the shielding material by photonuclear reactions were measured mostly with uranium fission-track dosimeters. The highest equivalent dose rate was found on the back of the emitter head; it is about 17 rem/h. The maximum value of the equivalent dose rate of neutrons on the patient's table amounts to 1,4 rem/h and is almost independent of the size of the diaphragm. The isodoses in the room depend largely on the shape of the irradiation room; they confirm the existence of a sufficient radioprotection.
Key concepts: Irradiation, Linear particle accelerator, Dosimeter, Neutron, Radiation, Thermoluminescence, Photon, Radiochemistry