Method performance verification of the clinical chemiluminescence immunoassay

Liang Wei-xion

Open publisher page 0 citations

Abstract

Objective To establish a method performance verification project and experimental method for the clinical chemiluminescence immunoassay.Methods Referring to CLSI evaluation protocols and pertinent literature,and by combining our actual works,we designed a verification procedure and experimental method.By Using these above,the precision,accuracy,analytical sensitivity,analytical measurement range,clinical reportable range and biotic interval of AFP on the Bayer Centaur 240 chemiluminescence immunoassay system were verificated.Results would be compared with the declaration of the manufacturer or desirable specifications derived from biologic variation.Results The results showed that the between-day inaccuracy on AFP levels at 77.4 ng/ml and 168.0 ng/ml was 5.70% and 4.84% respectively,these were consistent with manufacturer's inaccuracy claimed.The relative bias between the results measured for calibrator at four levels and target value was less 5.0%,and the relative bias between the results measured for EQA control sample at five levels and target value was-3.4% to 11.9%.Lower limit of detection was 1.04 ng/ml,lower slightly manufacturer's analytical sensitivity claimed.Biologic limit of detection was 2.65 ng/ml-3.53 ng/ml,functional sensitivity was 3.53 ng/ml.Analytical measurement range was 3.53-912.00 ng/ml,within manufacturer's liner range claimed.Clinical reportable range was 3.53-182 400.00 ng/ml.Reference interval was 0.6-7.7 ng/ml,within manufacturer' s claimed.Conclusions The main performances of the detection system are accorded with the declaration of the manufacturer.The performance verification procedure and experimental method of our research ars simple and practical,which has important significations for building medical laboratory and laboratory accreditation, improving quality of the chemiluminescence immunoassay.

About this research paper

What this paper is about

Objective To establish a method performance verification project and experimental method for the clinical chemiluminescence immunoassay.Methods Referring to CLSI evaluation protocols and pertinent literature,and by combining our actual works,we designed a verification procedure and experimental method.By Using these above,the precision,accuracy,analytical sensitivity,analytical measurement range,clinical reportable range and biotic interval of AFP on the Bayer Centaur 240 chemiluminescence immunoassay system were verificated.Results would be compared with the declaration of the manufacturer or desirable specifications derived from biologic variation.Results The results showed that the between-day inaccuracy on AFP levels at 77.4 ng/ml and 168.0 ng/ml was 5.70% and 4.84% respectively,these were consistent with manufacturer's inaccuracy claimed.The relative bias between the results measured for calibrator at four levels and target value was less 5.0%,and the relative bias between the results measured for EQA control sample at five levels and target value was-3.4% to 11.9%.Lower limit of detection was 1.04 ng/ml,lower slightly manufacturer's analytical sensitivity claimed.Biologic limit of detection was 2.65 ng/ml-3.53 ng/ml,functional sensitivity was 3.53 ng/ml.Analytical measurement range was 3.53-912.00 ng/ml,within manufacturer's liner range claimed.Clinical reportable range was 3.53-182 400.00 ng/ml.Reference interval was 0.6-7.7 ng/ml,within manufacturer' s claimed.Conclusions The main performances of the detection system are accorded with the declaration of the manufacturer.The performance verification procedure and experimental method of our research ars simple and practical,which has important significations for building medical laboratory and laboratory accreditation, improving quality of the chemiluminescence immunoassay.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To establish a method performance verification project and experimental method for the clinical chemiluminescence immunoassay.Methods Referring to CLSI evaluation protocols and pertinent literature,and by combining our actual works,we designed a verification procedure and experimental method.By Using these above,the precision,accuracy,analytical sensitivity,analytical measurement range,clinical reportable range and biotic interval of AFP on the Bayer Centaur 240 chemiluminescence immunoassay system were verificated.Results would be compared with the declaration of the manufacturer or desirable specifications derived from biologic variation.Results The results showed that the between-day inaccuracy on AFP levels at 77.4 ng/ml and 168.0 ng/ml was 5.70% and 4.84% respectively,these were consistent with manufacturer's inaccuracy claimed.The relative bias between the results measured for calibrator at four levels and target value was less 5.0%,and the relative bias between the results measured for EQA control sample at five levels and target value was-3.4% to 11.9%.Lower limit of detection was 1.04 ng/ml,lower slightly manufacturer's analytical sensitivity claimed.Biologic limit of detection was 2.65 ng/ml-3.53 ng/ml,functional sensitivity was 3.53 ng/ml.Analytical measurement range was 3.53-912.00 ng/ml,within manufacturer's liner range claimed.Clinical reportable range was 3.53-182 400.00 ng/ml.Reference interval was 0.6-7.7 ng/ml,within manufacturer' s claimed.Conclusions The main performances of the detection system are accorded with the declaration of the manufacturer.The performance verification procedure and experimental method of our research ars simple and practical,which has important significations for building medical laboratory and laboratory accreditation, improving quality of the chemiluminescence immunoassay.

Key concepts: Chemiluminescence, Detection limit, Immunoassay, Chromatography, Chemiluminescent immunoassay, Sensitivity (control systems), Declaration, Range (aeronautics)

Related papers

Back to paper searchBrowse research topicsOriginal source
Method performance verification of the clinical chemiluminescence immunoassay — Research Paper | ScholarLens