Accuracy and radiation exposure from image-guidance in Tomotherapy Hi-Art System

Jae-Uk Jang, Hyun-Soo Lim, Man-Seok Han, Yong Kyun Kim, Min-Cheol Jeon

Open publisher page 0 citations

Abstract

IGRT(Image Guided Radiation Therapy) in radiation therapy is a very useful technique in order to increase setup of patient and position reproducibility. Tomotherapy can increase accuracy of setup to take IGRT by MVCT, but it be for verified accuracy of Image guided, and MVCT occurs the exposure of patient. Through this study, IGRT accuracy of Tomotherapy is very accurate within 1.0mm. When MVCT using Tomotherapy phantom for QA, QC be taken, exposure dose is Fine(2mm Slice thickness) 3cGy, Normal(4mm Slice thickness) 1.5cGy, Corse(6mmSlice thickness) 1.0cGy. Measurement value of spatial resolution using AAPM CT performance phantom did't cause a big difference. As a result, ability of IGRT in Tomotherapy is very accurate. While obtaining image for IGRT, we should minimize expose range because patient's be exposed to radiation. We should make an effort to do accurate radiation therapy to minimize exposure of patient by selecting the appropriate thickness of MVCT depending on patient's body and treat area.

About this research paper

What this paper is about

IGRT(Image Guided Radiation Therapy) in radiation therapy is a very useful technique in order to increase setup of patient and position reproducibility. Tomotherapy can increase accuracy of setup to take IGRT by MVCT, but it be for verified accuracy of Image guided, and MVCT occurs the exposure of patient. Through this study, IGRT accuracy of Tomotherapy is very accurate within 1.0mm. When MVCT using Tomotherapy phantom for QA, QC be taken, exposure dose is Fine(2mm Slice thickness) 3cGy, Normal(4mm Slice thickness) 1.5cGy, Corse(6mmSlice thickness) 1.0cGy. Measurement value of spatial resolution using AAPM CT performance phantom did't cause a big difference. As a result, ability of IGRT in Tomotherapy is very accurate. While obtaining image for IGRT, we should minimize expose range because patient's be exposed to radiation. We should make an effort to do accurate radiation therapy to minimize exposure of patient by selecting the appropriate thickness of MVCT depending on patient's body and treat area.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

IGRT(Image Guided Radiation Therapy) in radiation therapy is a very useful technique in order to increase setup of patient and position reproducibility. Tomotherapy can increase accuracy of setup to take IGRT by MVCT, but it be for verified accuracy of Image guided, and MVCT occurs the exposure of patient. Through this study, IGRT accuracy of Tomotherapy is very accurate within 1.0mm. When MVCT using Tomotherapy phantom for QA, QC be taken, exposure dose is Fine(2mm Slice thickness) 3cGy, Normal(4mm Slice thickness) 1.5cGy, Corse(6mmSlice thickness) 1.0cGy. Measurement value of spatial resolution using AAPM CT performance phantom did't cause a big difference. As a result, ability of IGRT in Tomotherapy is very accurate. While obtaining image for IGRT, we should minimize expose range because patient's be exposed to radiation. We should make an effort to do accurate radiation therapy to minimize exposure of patient by selecting the appropriate thickness of MVCT depending on patient's body and treat area.

Key concepts: Tomotherapy, Image-guided radiation therapy, Imaging phantom, Medical physics, Medicine, Nuclear medicine, Reproducibility, Radiation dose

Related papers

Back to paper searchBrowse research topicsOriginal source
Accuracy and radiation exposure from image-guidance in Tomotherapy Hi-Art System — Research Paper | ScholarLens