Germicidal Efficacy of YKX/Y-Type Medical Air Disinfector:An Experimental Observation
Quanxue Lan
Abstract
Quanxue Lan
Abstract
OBJECTIVE To study the disinfection effect of YKX/Y medical air disinfector. METHODS The level of noise,the volume of circulating ventilation,UV intensity and leak of O 3 after the operation were determined. In the airtight room with controlled temperature and humidity,the concentration of O 3 was determined at different disinfection time. Two rooms with the same characteristics were selected to conduct the spot sterilization and spot-simulative sterilization experiment. RESULTS The results showed that under normal temperature there was no O 3 determined,and no UV leak after disinfection for 30,60,120 and 240 min. The volume of circulating ventilation was 1 510.2 m 3/h and the highest level of noise,63 dB(A). In 20 m 3 room,under the conditions of 20-22℃ and 65-70% relative humidity,disinfection for 30 min could eliminate 99.93% of Staphylococcus aureus ,And in 120 m 3 room,under the condition of 22-25℃ and 60-68% relative humidity,disinfection for 2 h could eliminate 93.13% of natural bacteria,with 67 CFU/m 3 bacteria remaining in air after disinfection. CONCLUSIONS YKX/Y medical air disinfector is appropriate to conduct routine air disinfection in the wards and clinic in the hospital with or without workers in it.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
OBJECTIVE To study the disinfection effect of YKX/Y medical air disinfector. METHODS The level of noise,the volume of circulating ventilation,UV intensity and leak of O 3 after the operation were determined. In the airtight room with controlled temperature and humidity,the concentration of O 3 was determined at different disinfection time. Two rooms with the same characteristics were selected to conduct the spot sterilization and spot-simulative sterilization experiment. RESULTS The results showed that under normal temperature there was no O 3 determined,and no UV leak after disinfection for 30,60,120 and 240 min. The volume of circulating ventilation was 1 510.2 m 3/h and the highest level of noise,63 dB(A). In 20 m 3 room,under the conditions of 20-22℃ and 65-70% relative humidity,disinfection for 30 min could eliminate 99.93% of Staphylococcus aureus ,And in 120 m 3 room,under the condition of 22-25℃ and 60-68% relative humidity,disinfection for 2 h could eliminate 93.13% of natural bacteria,with 67 CFU/m 3 bacteria remaining in air after disinfection. CONCLUSIONS YKX/Y medical air disinfector is appropriate to conduct routine air disinfection in the wards and clinic in the hospital with or without workers in it.
Key concepts: Sterilization (economics), Relative humidity, Humidity, Medicine, Toxicology, Biology, Economics, Foreign exchange market