Design and Application of Gas Flowmeter On-line Calibration System

LV Yun-pen

Open publisher page 0 citations

Abstract

Considering the high-cost of off-line calibration for gas flowmeter and the high accuracy and stability of ultrasonic flowmeter,a set of on-line calibration standard device for gas flowmeter was developed based on the working principle of time-difference ultrasonic flowmeter. Experiments proved that it can realize data acquisition,display and data processing of the standard ultrasonic flowmeter and checked flowmeter,and analyzed the factors which influence the accuracy of detection. It provides an effective protection for the measurement and quantity traceability of gas flowmeter.

About this research paper

What this paper is about

Considering the high-cost of off-line calibration for gas flowmeter and the high accuracy and stability of ultrasonic flowmeter,a set of on-line calibration standard device for gas flowmeter was developed based on the working principle of time-difference ultrasonic flowmeter. Experiments proved that it can realize data acquisition,display and data processing of the standard ultrasonic flowmeter and checked flowmeter,and analyzed the factors which influence the accuracy of detection. It provides an effective protection for the measurement and quantity traceability of gas flowmeter.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Considering the high-cost of off-line calibration for gas flowmeter and the high accuracy and stability of ultrasonic flowmeter,a set of on-line calibration standard device for gas flowmeter was developed based on the working principle of time-difference ultrasonic flowmeter. Experiments proved that it can realize data acquisition,display and data processing of the standard ultrasonic flowmeter and checked flowmeter,and analyzed the factors which influence the accuracy of detection. It provides an effective protection for the measurement and quantity traceability of gas flowmeter.

Key concepts: Flow measurement, Ultrasonic flow meter, Calibration, Ultrasonic sensor, Traceability, Line (geometry), Measuring principle, Engineering