Verification of GGT Detection Results with Two Measurement Systems by IFCC Reference Method

Qi Chen

Open publisher page 0 citations

Abstract

20 fresh,single human serum samples with different concentrations,ERM-AD452 and RELA samples were simultaneously measured by using A,B two factories measurement systems and IFCC reference method.The results measured with two methods were evaluated by software MVS and Matlab.The results showed that the GGT concentration in serum samples measured with A and B methods were obviously different.The measured results with method A was consistent with IFCC reference method,and the regression equation between method A and IFCC reference method was Y=0.995X- 0.417,and the prediction bias value was 0.359.The measured results with method B were inconsistent with IFCC reference method,the regression equation between method B and IFCC reference method was Y=0.827X- 0.566,and the prediction bias value was 17.258.In view of obvious difference of results with different measurement systems,the laboratories should establish a method to evaluate routine measurement results with different methods.

About this research paper

What this paper is about

20 fresh,single human serum samples with different concentrations,ERM-AD452 and RELA samples were simultaneously measured by using A,B two factories measurement systems and IFCC reference method.The results measured with two methods were evaluated by software MVS and Matlab.The results showed that the GGT concentration in serum samples measured with A and B methods were obviously different.The measured results with method A was consistent with IFCC reference method,and the regression equation between method A and IFCC reference method was Y=0.995X- 0.417,and the prediction bias value was 0.359.The measured results with method B were inconsistent with IFCC reference method,the regression equation between method B and IFCC reference method was Y=0.827X- 0.566,and the prediction bias value was 17.258.In view of obvious difference of results with different measurement systems,the laboratories should establish a method to evaluate routine measurement results with different methods.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

20 fresh,single human serum samples with different concentrations,ERM-AD452 and RELA samples were simultaneously measured by using A,B two factories measurement systems and IFCC reference method.The results measured with two methods were evaluated by software MVS and Matlab.The results showed that the GGT concentration in serum samples measured with A and B methods were obviously different.The measured results with method A was consistent with IFCC reference method,and the regression equation between method A and IFCC reference method was Y=0.995X- 0.417,and the prediction bias value was 0.359.The measured results with method B were inconsistent with IFCC reference method,the regression equation between method B and IFCC reference method was Y=0.827X- 0.566,and the prediction bias value was 17.258.In view of obvious difference of results with different measurement systems,the laboratories should establish a method to evaluate routine measurement results with different methods.

Key concepts: Medicine, Reference values, Linear regression, Statistics, Mathematics, Internal medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
Verification of GGT Detection Results with Two Measurement Systems by IFCC Reference Method — Research Paper | ScholarLens