Determination of Residual Organic Solvents in Arbidol by Capillary GC
Kang Xue
Abstract
Kang Xue
Abstract
A capillary GC method with FID detector for determination of residual organic solvents——methanol, ethanol, acetone, ethyl ether and petroleum ether (60~90 ℃) in arbidol was established. An OV 101 capillary column (0.22 mm×30 m) was used with procedural column temperature control. Nitrogen gas was used as carrier gas. The detection limits were 5,5,3,3 and 10 ng, respectively. The calibration curves of the five solvents were linear ( r 0.997). The average recoveries were 99.4%, 101.6% ,101.1%,101.9% and 97.1%, respectively.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A capillary GC method with FID detector for determination of residual organic solvents——methanol, ethanol, acetone, ethyl ether and petroleum ether (60~90 ℃) in arbidol was established. An OV 101 capillary column (0.22 mm×30 m) was used with procedural column temperature control. Nitrogen gas was used as carrier gas. The detection limits were 5,5,3,3 and 10 ng, respectively. The calibration curves of the five solvents were linear ( r 0.997). The average recoveries were 99.4%, 101.6% ,101.1%,101.9% and 97.1%, respectively.
Key concepts: Chemistry, Acetone, Chromatography, Methanol, Capillary action, Gas chromatography, Calibration curve, Detection limit