2007•Laboratory MedicineRequires access

The coincidence of the results from Beckman automatic biochemical analyzer using Kemei reagent and Beckman reagent

Luo Hu

Open publisher page 0 citations

Abstract

Objective To study the coincidence of the results from Beckman LX20 automatic biochemical analyzer using Kemei reagent and Beckman reagent. Methods Fresh mixed patient sera used as temporary calibrator was determined by Beckman reagent and Kemei reagent.The value of RANDOX calibrator was gotten from the calibrated Kemei reagent system and applied in it.Glucose(Glu),triglyceride(TG),cholesterol(Chol),urea,total protein(TP),albumin(Alb) and uric acid(UA) were detected in 40 samples with the two reagents at the same time. Results In comparison with the results using Beckman reagent,the results using Kemei reagent was lower for UA,the other six items coincided with the Beckman reagent system.Conclusions The fresh mixed patient sera can transmit the traceability of Beckman regeat system to the Kemei reagent.

About this research paper

What this paper is about

Objective To study the coincidence of the results from Beckman LX20 automatic biochemical analyzer using Kemei reagent and Beckman reagent. Methods Fresh mixed patient sera used as temporary calibrator was determined by Beckman reagent and Kemei reagent.The value of RANDOX calibrator was gotten from the calibrated Kemei reagent system and applied in it.Glucose(Glu),triglyceride(TG),cholesterol(Chol),urea,total protein(TP),albumin(Alb) and uric acid(UA) were detected in 40 samples with the two reagents at the same time. Results In comparison with the results using Beckman reagent,the results using Kemei reagent was lower for UA,the other six items coincided with the Beckman reagent system.Conclusions The fresh mixed patient sera can transmit the traceability of Beckman regeat system to the Kemei reagent.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To study the coincidence of the results from Beckman LX20 automatic biochemical analyzer using Kemei reagent and Beckman reagent. Methods Fresh mixed patient sera used as temporary calibrator was determined by Beckman reagent and Kemei reagent.The value of RANDOX calibrator was gotten from the calibrated Kemei reagent system and applied in it.Glucose(Glu),triglyceride(TG),cholesterol(Chol),urea,total protein(TP),albumin(Alb) and uric acid(UA) were detected in 40 samples with the two reagents at the same time. Results In comparison with the results using Beckman reagent,the results using Kemei reagent was lower for UA,the other six items coincided with the Beckman reagent system.Conclusions The fresh mixed patient sera can transmit the traceability of Beckman regeat system to the Kemei reagent.

Key concepts: Reagent, Chemistry, Chromatography, Uric acid, Biochemistry, Organic chemistry