Calibration of Relative Detection Efficiency for Neutron Long Counter
Lin Mei
Abstract
Lin Mei
Abstract
The relative detection efficiency of neutron long counter was calibrated with 241Am-Be neutron source.Energy conversion coefficient between 5.1 MeV and 14 MeV was also determined by the associated particle method.At the same time,the relative detection efficiency and energy conversion coefficient were calculated with Monte-Carlo method.In present work,the uncertainty of relative detection efficiency from experiment and calculation is 10.8% and 2.3%,respectively.Both are coincident in the range of uncertainty.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The relative detection efficiency of neutron long counter was calibrated with 241Am-Be neutron source.Energy conversion coefficient between 5.1 MeV and 14 MeV was also determined by the associated particle method.At the same time,the relative detection efficiency and energy conversion coefficient were calculated with Monte-Carlo method.In present work,the uncertainty of relative detection efficiency from experiment and calculation is 10.8% and 2.3%,respectively.Both are coincident in the range of uncertainty.