Evaluation of Sysmex XE-2100 Hematology Analyzer
Yang Zhang-lan
Abstract
Yang Zhang-lan
Abstract
Objective To evaluate primarily SYSMEX XE-2100 Hematology Analyzer.Methods Mensurated the precision,linearity,relativity and Carryover in SYSMEX XE-2100 by the samples from healthy people and outpatients.Some of the result was compared with sysmex NE-1500 Hematology Analyzer's.Results Precision,accuracy,reproducibility and stability are outstanding.Linearity is wide and Carryover is low in SYSMEX XE-2100 Hematology Analyzer.Conclusion SYSMEX XE-2100 Hematology Analyzer is very suit for applying in outpatient department of clinical laboratory.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To evaluate primarily SYSMEX XE-2100 Hematology Analyzer.Methods Mensurated the precision,linearity,relativity and Carryover in SYSMEX XE-2100 by the samples from healthy people and outpatients.Some of the result was compared with sysmex NE-1500 Hematology Analyzer's.Results Precision,accuracy,reproducibility and stability are outstanding.Linearity is wide and Carryover is low in SYSMEX XE-2100 Hematology Analyzer.Conclusion SYSMEX XE-2100 Hematology Analyzer is very suit for applying in outpatient department of clinical laboratory.
Key concepts: Hematology analyzer, Medicine, Spectrum analyzer, Reproducibility, Hematology, Internal medicine, Medical physics, Nuclear medicine