The clinical application value of CARE kV combined with CARE Dose 4D techniques in the orbital CT scan

Zeng Xian-chu

Open publisher page 0 citations

Abstract

Objective:To explore the clinical application value of CARE Dose 4D(tube current regulation technique)combined with CARE kV(the smart kV scanning technology)in orbital CT scan,expecting to reduce the radiation dose.Methods:70 patients underwent plain orbital CT scan,they were randomly divided into two groups according to chronological sequence.In group A(n=35 cases),applied CARE Dose 4D technique was used only;In group B(n=35 cases),applied both CARE Dose 4D and CARE kV technique was used.Comparison of radiation dose and image quality(based on average CT values,Noise SD,SNR,CNR and quality score)between the two groups were made.Results:The difference of radiation dose between the two groups was statistically significant(P0.05);ED of group B was less than those of group A by around 16.9%.There were no significant differences between the two groups in average CT values,SD,SNR and CNR(P0.05);Also the image quality score and lesion detection rate demonstrated no significant difference among the two groups.Conclusion:On the premise of obtaining good quality of image for diagnosis,CARE kV combined with CARE Dose 4D technology can be applied in orbital CT scan,which will reduce about 15.9% radiation dose.

About this research paper

What this paper is about

Objective:To explore the clinical application value of CARE Dose 4D(tube current regulation technique)combined with CARE kV(the smart kV scanning technology)in orbital CT scan,expecting to reduce the radiation dose.Methods:70 patients underwent plain orbital CT scan,they were randomly divided into two groups according to chronological sequence.In group A(n=35 cases),applied CARE Dose 4D technique was used only;In group B(n=35 cases),applied both CARE Dose 4D and CARE kV technique was used.Comparison of radiation dose and image quality(based on average CT values,Noise SD,SNR,CNR and quality score)between the two groups were made.Results:The difference of radiation dose between the two groups was statistically significant(P0.05);ED of group B was less than those of group A by around 16.9%.There were no significant differences between the two groups in average CT values,SD,SNR and CNR(P0.05);Also the image quality score and lesion detection rate demonstrated no significant difference among the two groups.Conclusion:On the premise of obtaining good quality of image for diagnosis,CARE kV combined with CARE Dose 4D technology can be applied in orbital CT scan,which will reduce about 15.9% radiation dose.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective:To explore the clinical application value of CARE Dose 4D(tube current regulation technique)combined with CARE kV(the smart kV scanning technology)in orbital CT scan,expecting to reduce the radiation dose.Methods:70 patients underwent plain orbital CT scan,they were randomly divided into two groups according to chronological sequence.In group A(n=35 cases),applied CARE Dose 4D technique was used only;In group B(n=35 cases),applied both CARE Dose 4D and CARE kV technique was used.Comparison of radiation dose and image quality(based on average CT values,Noise SD,SNR,CNR and quality score)between the two groups were made.Results:The difference of radiation dose between the two groups was statistically significant(P0.05);ED of group B was less than those of group A by around 16.9%.There were no significant differences between the two groups in average CT values,SD,SNR and CNR(P0.05);Also the image quality score and lesion detection rate demonstrated no significant difference among the two groups.Conclusion:On the premise of obtaining good quality of image for diagnosis,CARE kV combined with CARE Dose 4D technology can be applied in orbital CT scan,which will reduce about 15.9% radiation dose.

Key concepts: Medicine, Radiation dose, Image quality, Nuclear medicine, Computed tomography, Effective dose (radiation), Radiology, Image noise

Related papers

Back to paper searchBrowse research topicsOriginal source
The clinical application value of CARE kV combined with CARE Dose 4D techniques in the orbital CT scan — Research Paper | ScholarLens