Determination of Cadmium in Sewage by Flame Emission Atomic Absorption Spectrometry
Jun Xie
Abstract
Jun Xie
Abstract
s:The sewage sample was digested with nitric acid-perchloric acid, the cadmium in which was determined by flame emission atomic absorption spectrometry. Under the optimum experimental conditions, the linear range of the method was 0~1.0 mg/L, the detection limit was 0.05 mg/L, the relative standard deviation was less than 2.58% and the recovery was 97%~102.0%.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
s:The sewage sample was digested with nitric acid-perchloric acid, the cadmium in which was determined by flame emission atomic absorption spectrometry. Under the optimum experimental conditions, the linear range of the method was 0~1.0 mg/L, the detection limit was 0.05 mg/L, the relative standard deviation was less than 2.58% and the recovery was 97%~102.0%.