Application of Safe-Cut~(TM) balloon in coronary artery in-stent restenosis
Jiang Jin-fa
Abstract
Jiang Jin-fa
Abstract
Objective To assess the safety and therapeutical efficiency of dual-wire balloon using in treating coronary artery in-stent restenosis.Methods Eighty-two lesions out of 70 patients with more than 70% restenosis were included after coronary angiography.Dual-wire balloon angioplasty was performed in these lesions.Quantitative coronary angiography(QCA)analysis was used to analyze dilatation efficiency and clinical follow-up visit was also performed.Results Success rate of dilatation was 100%.QCA analysis indicated that target minimum lumen diameter(MLD)significantly increased from(0.76±0.52)mm to(2.64±0.41)mm(P0.001) while diameter stenosis(DS)decreased significantly from 87.7%±8.4% to 8.6%±6.9%(P0.001).No cardiovascular events occurred during the follow-up visit.Conclusion Dual-wire balloon angioplasty is effective and safe for the treatment of in-stent restenosis.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To assess the safety and therapeutical efficiency of dual-wire balloon using in treating coronary artery in-stent restenosis.Methods Eighty-two lesions out of 70 patients with more than 70% restenosis were included after coronary angiography.Dual-wire balloon angioplasty was performed in these lesions.Quantitative coronary angiography(QCA)analysis was used to analyze dilatation efficiency and clinical follow-up visit was also performed.Results Success rate of dilatation was 100%.QCA analysis indicated that target minimum lumen diameter(MLD)significantly increased from(0.76±0.52)mm to(2.64±0.41)mm(P0.001) while diameter stenosis(DS)decreased significantly from 87.7%±8.4% to 8.6%±6.9%(P0.001).No cardiovascular events occurred during the follow-up visit.Conclusion Dual-wire balloon angioplasty is effective and safe for the treatment of in-stent restenosis.
Key concepts: Restenosis, Medicine, Balloon, Angioplasty, Stent, Lumen (anatomy), Stenosis, Cardiology