Study on Flow Injection Spectrophotometric Determination of Nitrite in Water
Ting Ma
Abstract
Ting Ma
Abstract
A method was described for reversed phase flow injection spectrophotometric determination of trace nitrite in water. The linear range is 0.002 to 0.30 mg/L, nitrite nitrogcn with the detection limit of 0.0005 mg/L. This method has been applied to the determination of trace nitrite nitrogen in tap water. The results have been compared with those of the national standard method, indicating satisfactory precision and accuracy.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A method was described for reversed phase flow injection spectrophotometric determination of trace nitrite in water. The linear range is 0.002 to 0.30 mg/L, nitrite nitrogcn with the detection limit of 0.0005 mg/L. This method has been applied to the determination of trace nitrite nitrogen in tap water. The results have been compared with those of the national standard method, indicating satisfactory precision and accuracy.
Key concepts: Nitrite, Tap water, Detection limit, Chemistry, Chromatography, Linear range, TRACE (psycholinguistics), Flow injection analysis