Determination of Arsenic Cadmium Iron Mercury Lead in Selenium Dioxide by ICP-AES
Jinping Wang
Abstract
Jinping Wang
Abstract
In nitric acid medium,the elements of As,Cd,Fe,Hg,Pb in selenium dioxide by ICP-AES are directly determined.The result shows that nitric acid medium as 0.7 mol/L,selenium content as 7 mg/mL in solution,As,Cd,Fe,Hg,Pb detection limit are 0.060,0.036,0.056,0.021,0.068 μg/mL respectively,the recovery in standard is from 95.24% to 105.64%,relative standard deviation is from 0.0096% to 0.081%.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
In nitric acid medium,the elements of As,Cd,Fe,Hg,Pb in selenium dioxide by ICP-AES are directly determined.The result shows that nitric acid medium as 0.7 mol/L,selenium content as 7 mg/mL in solution,As,Cd,Fe,Hg,Pb detection limit are 0.060,0.036,0.056,0.021,0.068 μg/mL respectively,the recovery in standard is from 95.24% to 105.64%,relative standard deviation is from 0.0096% to 0.081%.
Key concepts: Selenium, Nitric acid, Arsenic, Mercury (programming language), Chemistry, Cadmium, Detection limit, Relative standard deviation