Determination of Perchlorate in Surface Water by High Performance Liquid Chromatography-Electrospray Ion Source-Tandem Mass Spectrometry

Xie Yong-hon

Open publisher page 0 citations

Abstract

A method was developed for determination of perchlorate in surface water based on high performance liquid chromatography-electrospray ion sourcetandem mass spectrometry( HPLC-ESI-MS / MS). The anion exchange column of Dionex IonPac AG20 was used as the analytical column with the weakly alkaline mobile phase of 0. 056% ammonia /5 mmol / L Ammonium acetate and the flow rate of 1. 0 mL / min.-ESI MS / MS was used as the detector and the perhclorate was detected under multiple reaction monitoring( MRM) mode. The detection limit of the developed method was 0. 043 μg / L and the linear range was 0. 2-50 μg / L with the linear correlation coefficient of 0. 999 4,the relative standard deviation for about 2,6,30 μg / L sample were 2. 47%,4. 55%,0. 49%,respectively. Content recovery percentages ranging from 80% to 109% were obtained and the analysis results were of agreement with that analyzed by US EPA method314. 0.

About this research paper

What this paper is about

A method was developed for determination of perchlorate in surface water based on high performance liquid chromatography-electrospray ion sourcetandem mass spectrometry( HPLC-ESI-MS / MS). The anion exchange column of Dionex IonPac AG20 was used as the analytical column with the weakly alkaline mobile phase of 0. 056% ammonia /5 mmol / L Ammonium acetate and the flow rate of 1. 0 mL / min.-ESI MS / MS was used as the detector and the perhclorate was detected under multiple reaction monitoring( MRM) mode. The detection limit of the developed method was 0. 043 μg / L and the linear range was 0. 2-50 μg / L with the linear correlation coefficient of 0. 999 4,the relative standard deviation for about 2,6,30 μg / L sample were 2. 47%,4. 55%,0. 49%,respectively. Content recovery percentages ranging from 80% to 109% were obtained and the analysis results were of agreement with that analyzed by US EPA method314. 0.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

A method was developed for determination of perchlorate in surface water based on high performance liquid chromatography-electrospray ion sourcetandem mass spectrometry( HPLC-ESI-MS / MS). The anion exchange column of Dionex IonPac AG20 was used as the analytical column with the weakly alkaline mobile phase of 0. 056% ammonia /5 mmol / L Ammonium acetate and the flow rate of 1. 0 mL / min.-ESI MS / MS was used as the detector and the perhclorate was detected under multiple reaction monitoring( MRM) mode. The detection limit of the developed method was 0. 043 μg / L and the linear range was 0. 2-50 μg / L with the linear correlation coefficient of 0. 999 4,the relative standard deviation for about 2,6,30 μg / L sample were 2. 47%,4. 55%,0. 49%,respectively. Content recovery percentages ranging from 80% to 109% were obtained and the analysis results were of agreement with that analyzed by US EPA method314. 0.

Key concepts: Chemistry, Chromatography, Detection limit, Electrospray, Analytical Chemistry (journal), Selected reaction monitoring, Mass spectrometry, Ion chromatography

Related papers

Back to paper searchBrowse research topicsOriginal source
Determination of Perchlorate in Surface Water by High Performance Liquid Chromatography-Electrospray Ion Source-Tandem Mass Spectrometry — Research Paper | ScholarLens