Evaluation of Uncertainty on Standard Device of Two Pressure Humidity Generator
Changqing Ren
Abstract
Changqing Ren
Abstract
In the article we introduce the specification of the two pressure humidity generator, measurement method, uncertainty component and its evaluation. The measurement results and uncertainty analysis shows that in the dew point range of -25℃- +25℃the uncertainty less than 0. 04℃. It meets the dissemination requirements of humidity value in this range.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
In the article we introduce the specification of the two pressure humidity generator, measurement method, uncertainty component and its evaluation. The measurement results and uncertainty analysis shows that in the dew point range of -25℃- +25℃the uncertainty less than 0. 04℃. It meets the dissemination requirements of humidity value in this range.
Key concepts: Humidity, Dew point, Measurement uncertainty, Generator (circuit theory), Dew, Range (aeronautics), Uncertainty analysis, Environmental science