Methodological evaluation of chemiluminescence microparticle immunoassay for determining HE4

Liang Yong-yuan

Open publisher page 0 citations

Abstract

Objective To evaluate the performance methodology of Chemiluminescence microparticle immunoassay(CMIA) for determining HE4 using Abbott Architect i2000SR Chemiluminescence analyzer.Methods According to a method by NCCLS recommended,we designed the evaluation program of quantitative detection methods.The within-run precision、between-run precision、linear range、carry-over and biological reference interval of HE4 were determined.Results The coefficient of variation(CV) of within-run precision of low and high value pooled serum were respectively 1.80% and 2.34%.The CV of between-run precision were respectively 2.03% and 2.70%.The linear range was from 19.8 pmol/L to 1 573.5 pmol/L.The 1carry-over was-0.02%.The biological reference interval was consistent with manufactures which is from 0.0 pmol/L to 140.0 pmol/L.Conclusion The within-run precision,between-run precision,linear range and carry-over of Abbott Architect i2000SR for determining HE4 meet the requirements of clinical use.The biological reference interval was consistent with that provided by the manufacturer.

About this research paper

What this paper is about

Objective To evaluate the performance methodology of Chemiluminescence microparticle immunoassay(CMIA) for determining HE4 using Abbott Architect i2000SR Chemiluminescence analyzer.Methods According to a method by NCCLS recommended,we designed the evaluation program of quantitative detection methods.The within-run precision、between-run precision、linear range、carry-over and biological reference interval of HE4 were determined.Results The coefficient of variation(CV) of within-run precision of low and high value pooled serum were respectively 1.80% and 2.34%.The CV of between-run precision were respectively 2.03% and 2.70%.The linear range was from 19.8 pmol/L to 1 573.5 pmol/L.The 1carry-over was-0.02%.The biological reference interval was consistent with manufactures which is from 0.0 pmol/L to 140.0 pmol/L.Conclusion The within-run precision,between-run precision,linear range and carry-over of Abbott Architect i2000SR for determining HE4 meet the requirements of clinical use.The biological reference interval was consistent with that provided by the manufacturer.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To evaluate the performance methodology of Chemiluminescence microparticle immunoassay(CMIA) for determining HE4 using Abbott Architect i2000SR Chemiluminescence analyzer.Methods According to a method by NCCLS recommended,we designed the evaluation program of quantitative detection methods.The within-run precision、between-run precision、linear range、carry-over and biological reference interval of HE4 were determined.Results The coefficient of variation(CV) of within-run precision of low and high value pooled serum were respectively 1.80% and 2.34%.The CV of between-run precision were respectively 2.03% and 2.70%.The linear range was from 19.8 pmol/L to 1 573.5 pmol/L.The 1carry-over was-0.02%.The biological reference interval was consistent with manufactures which is from 0.0 pmol/L to 140.0 pmol/L.Conclusion The within-run precision,between-run precision,linear range and carry-over of Abbott Architect i2000SR for determining HE4 meet the requirements of clinical use.The biological reference interval was consistent with that provided by the manufacturer.

Key concepts: Chemiluminescence, Coefficient of variation, Chemiluminescent immunoassay, Spectrum analyzer, Immunoassay, Accuracy and precision, Linear relationship, Interval (graph theory)

Related papers

Back to paper searchBrowse research topicsOriginal source
Methodological evaluation of chemiluminescence microparticle immunoassay for determining HE4 — Research Paper | ScholarLens