Measurement and evaluation of personal radiation dose in cyclotron center and PET-CT department

YU Xi-rong

Open publisher page 0 citations

Abstract

Objective To measure and evaluate the personal radiation dose for medical staff in cyclotron center and PET-CT department and to offer the reference data for clinical radiation protection.Methods γ ray radiation dose rate was analyzed in 18 medical staff members during each medical procedure by using X and γ radiation detector.The estimation of annual exposure dose of 18 medical staff at different working site was based on the caculation of doserate(d),work time(t) and workload(w).Results When the workloads were 1 000,4 000,7 000 and 10 000 per year,the annual doses per procedure were: 0.27 mSv for quality control;0.02 to 0.20 mSv for dispensation;0.06 to 0.58 mSv for carrying the compound;2.49 to 24.9 mSv for injection and 1.19 to 11.91 mSv for setting the patients.Conclusion The staff receives higher radiation dose during dispensation and injection.When the workload is 4 000 and 10 000 per year,the dose received by hands and body would respectively exceed the national radiation safety standards(GBSS) on occupational annual dose limits.

About this research paper

What this paper is about

Objective To measure and evaluate the personal radiation dose for medical staff in cyclotron center and PET-CT department and to offer the reference data for clinical radiation protection.Methods γ ray radiation dose rate was analyzed in 18 medical staff members during each medical procedure by using X and γ radiation detector.The estimation of annual exposure dose of 18 medical staff at different working site was based on the caculation of doserate(d),work time(t) and workload(w).Results When the workloads were 1 000,4 000,7 000 and 10 000 per year,the annual doses per procedure were: 0.27 mSv for quality control;0.02 to 0.20 mSv for dispensation;0.06 to 0.58 mSv for carrying the compound;2.49 to 24.9 mSv for injection and 1.19 to 11.91 mSv for setting the patients.Conclusion The staff receives higher radiation dose during dispensation and injection.When the workload is 4 000 and 10 000 per year,the dose received by hands and body would respectively exceed the national radiation safety standards(GBSS) on occupational annual dose limits.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To measure and evaluate the personal radiation dose for medical staff in cyclotron center and PET-CT department and to offer the reference data for clinical radiation protection.Methods γ ray radiation dose rate was analyzed in 18 medical staff members during each medical procedure by using X and γ radiation detector.The estimation of annual exposure dose of 18 medical staff at different working site was based on the caculation of doserate(d),work time(t) and workload(w).Results When the workloads were 1 000,4 000,7 000 and 10 000 per year,the annual doses per procedure were: 0.27 mSv for quality control;0.02 to 0.20 mSv for dispensation;0.06 to 0.58 mSv for carrying the compound;2.49 to 24.9 mSv for injection and 1.19 to 11.91 mSv for setting the patients.Conclusion The staff receives higher radiation dose during dispensation and injection.When the workload is 4 000 and 10 000 per year,the dose received by hands and body would respectively exceed the national radiation safety standards(GBSS) on occupational annual dose limits.

Key concepts: Workload, Radiation protection, Medicine, Nuclear medicine, Effective dose (radiation), Medical physics, Equivalent dose, Radiation dose