Using real-time three-dimensional echocardiography to quantify in vivo ventricular mass

Guoying Huang

Open publisher page 0 citations

Abstract

Purpose This in vivo experimental study was designed to evaluate the feasibility and accuracy of real-time three-dimensional echocardiography(RT-3DE) for quantifying ventricular mass. Methods The hearts of twelve normal experimental pigs [weight(13-23.5) kg,mean(20.17±(2.79)) kg] were examined by RT-3DE and two-dimensional echocardiography(2DE) with Philips/Sonos 7500 ultrasound system.The ventricular wall mass was outlined and calculated by RT-3DE 8-plane method with the 4D Cardio-View RT of Tomtec image system and 2DE Simpson method.The true value of porcine ventricular wall mass weighed by the scales was served as a reference standard. Results Full volume imaging in all porcine hearts was well displayed by RT-3DE.The measurements of left ventricular end-diastolic(62.37±5.63) g and end-systolic(61.26±7.89) g wall mass with RT-3DE were close to the true value of porcine left ventricular wall mass(62.99±9.67) g with well correlation(r=0.88,r=0.86).The measurements of right ventricular end-diastolic(19.66±2.95) g and end-systolic(20.18±2.87) g free wall mass with RT-3DE were close to the true value of porcine right ventricular free wall mass(20.56±3.22) g with well correlation(r=0.88,r=0.87).The measurements with 2DE were underestimated and poorly correlated with the true value. Conclusions RT-3DE is a valuable technique to quantify in vivo ventricular wall mass.

About this research paper

What this paper is about

Purpose This in vivo experimental study was designed to evaluate the feasibility and accuracy of real-time three-dimensional echocardiography(RT-3DE) for quantifying ventricular mass. Methods The hearts of twelve normal experimental pigs [weight(13-23.5) kg,mean(20.17±(2.79)) kg] were examined by RT-3DE and two-dimensional echocardiography(2DE) with Philips/Sonos 7500 ultrasound system.The ventricular wall mass was outlined and calculated by RT-3DE 8-plane method with the 4D Cardio-View RT of Tomtec image system and 2DE Simpson method.The true value of porcine ventricular wall mass weighed by the scales was served as a reference standard. Results Full volume imaging in all porcine hearts was well displayed by RT-3DE.The measurements of left ventricular end-diastolic(62.37±5.63) g and end-systolic(61.26±7.89) g wall mass with RT-3DE were close to the true value of porcine left ventricular wall mass(62.99±9.67) g with well correlation(r=0.88,r=0.86).The measurements of right ventricular end-diastolic(19.66±2.95) g and end-systolic(20.18±2.87) g free wall mass with RT-3DE were close to the true value of porcine right ventricular free wall mass(20.56±3.22) g with well correlation(r=0.88,r=0.87).The measurements with 2DE were underestimated and poorly correlated with the true value. Conclusions RT-3DE is a valuable technique to quantify in vivo ventricular wall mass.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Purpose This in vivo experimental study was designed to evaluate the feasibility and accuracy of real-time three-dimensional echocardiography(RT-3DE) for quantifying ventricular mass. Methods The hearts of twelve normal experimental pigs [weight(13-23.5) kg,mean(20.17±(2.79)) kg] were examined by RT-3DE and two-dimensional echocardiography(2DE) with Philips/Sonos 7500 ultrasound system.The ventricular wall mass was outlined and calculated by RT-3DE 8-plane method with the 4D Cardio-View RT of Tomtec image system and 2DE Simpson method.The true value of porcine ventricular wall mass weighed by the scales was served as a reference standard. Results Full volume imaging in all porcine hearts was well displayed by RT-3DE.The measurements of left ventricular end-diastolic(62.37±5.63) g and end-systolic(61.26±7.89) g wall mass with RT-3DE were close to the true value of porcine left ventricular wall mass(62.99±9.67) g with well correlation(r=0.88,r=0.86).The measurements of right ventricular end-diastolic(19.66±2.95) g and end-systolic(20.18±2.87) g free wall mass with RT-3DE were close to the true value of porcine right ventricular free wall mass(20.56±3.22) g with well correlation(r=0.88,r=0.87).The measurements with 2DE were underestimated and poorly correlated with the true value. Conclusions RT-3DE is a valuable technique to quantify in vivo ventricular wall mass.

Key concepts: In vivo, Ventricular volume, Diastole, Medicine, Ultrasound, Cardiology, Internal medicine, Nuclear medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
Using real-time three-dimensional echocardiography to quantify in vivo ventricular mass — Research Paper | ScholarLens