Simultaneous determinination of seven amions in drinking water by ion chromatography
Zheng Jiang
Abstract
Zheng Jiang
Abstract
Objective:To develop a method with ion chromatography to simultaneously determine seven anions in drinking water.Methods: The samples were filtered through a 0.2 μm millipore filter and injected directly.The seven anions in samples were separated by an Ionpac AS19 analytical column(4 mm×250 mm) with Ionpac AG19 guard column and 34.0 mmol/L KOH as eluent.The flow rate was 1.0 ml/min.Anions were detected by suppressed electric conductivity detector.Results: This method exhibited excellent correlation(r0.9993) and good analytic precision(RSD3.0%).The recovery rates were in the range of 95.0%~103.7%.The detection limit were F-: 0.0015 mg/L;Cl-: 0.0010 mg/L;NO-2-N: 0.0019 mg/L;SO2-4: 0.0012 mg/L;Br-: 0.0018 mg/L;NO-3: 0.0013 mg/L;HPO2-4: 0.0023 mg/L.Conclusion: This method was convenient,fast and accurate with wide linear range and it can be applied for inspection of large numbers of water samples.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective:To develop a method with ion chromatography to simultaneously determine seven anions in drinking water.Methods: The samples were filtered through a 0.2 μm millipore filter and injected directly.The seven anions in samples were separated by an Ionpac AS19 analytical column(4 mm×250 mm) with Ionpac AG19 guard column and 34.0 mmol/L KOH as eluent.The flow rate was 1.0 ml/min.Anions were detected by suppressed electric conductivity detector.Results: This method exhibited excellent correlation(r0.9993) and good analytic precision(RSD3.0%).The recovery rates were in the range of 95.0%~103.7%.The detection limit were F-: 0.0015 mg/L;Cl-: 0.0010 mg/L;NO-2-N: 0.0019 mg/L;SO2-4: 0.0012 mg/L;Br-: 0.0018 mg/L;NO-3: 0.0013 mg/L;HPO2-4: 0.0023 mg/L.Conclusion: This method was convenient,fast and accurate with wide linear range and it can be applied for inspection of large numbers of water samples.
Key concepts: Ion chromatography, Chemistry, Detection limit, Chromatography, Column chromatography, Thermal conductivity detector, Conductivity, Ion