Catheterization of the Heart in Man with Use of a Flow-Directed Balloon-Tipped Catheter
Harold J.C. Swan, William Ganz, James S. Forrester, Harold S. Marcus, George A. Diamond, David Chonette
Abstract
Harold J.C. Swan, William Ganz, James S. Forrester, Harold S. Marcus, George A. Diamond, David Chonette
Abstract
Pressures in the right side of the heart and pulmonary capillary wedge can be obtained by cardiac catheterization without the aid of fluoroscopy. A No. 5 Fr double-lumen catheter with a balloon just proximal to the tip is inserted into the right atrium under pressure monitoring. The balloon is then inflated with 0.8 ml of air. The balloon is carried by blood flow through the right side of the heart into the smaller radicles of the pulmonary artery. In this position when the balloon is inflated wedge pressure is obtained. The average time for passage of the catheter from the right atrium to the pulmonary artery was 35 seconds in the first 100 passages. The frequency of premature beats was minimal, and no other arrhythmias occurred.
OpenAlex reports 2146 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Pressures in the right side of the heart and pulmonary capillary wedge can be obtained by cardiac catheterization without the aid of fluoroscopy. A No. 5 Fr double-lumen catheter with a balloon just proximal to the tip is inserted into the right atrium under pressure monitoring. The balloon is then inflated with 0.8 ml of air. The balloon is carried by blood flow through the right side of the heart into the smaller radicles of the pulmonary artery. In this position when the balloon is inflated wedge pressure is obtained. The average time for passage of the catheter from the right atrium to the pulmonary artery was 35 seconds in the first 100 passages. The frequency of premature beats was minimal, and no other arrhythmias occurred.
Key concepts: Medicine, Pulmonary wedge pressure, Balloon, Catheter, Pulmonary artery, Cardiac catheterization, Heart catheterization, Cardiology