1987Laboratory MedicineRequires access

A Multiparameter Automated Hematology Analyzer: The Technicon H-6000

Pamela Bollinger, Benjamin Drewinko, Carrie D. Brailas, Sheila Hickey, Dennis A. Johnston, José M. Trujillo

Open publisher page 1 citations

Abstract

The performance of the Technicon H-6000 was evaluated in the computerized, high-volume hematology laboratory at M.D. Anderson Hospital and Tumor Institute, Houston, Tex, and compared with that of reference instrumentation used for routine patient care. The within-run precision, linearity, and lack of carryover was excellent, and the day-to-day precision was satisfactory for the entire range of tested parameters (WBC, RBC, HGB, HCT, MCV, RDW, and PLT). Coefficients of correlation for these parameters were 0.98 or greater, with the exception of the MCV (r = 0.93). Special emphasis was placed on the correlation study of WBCs and PLTs at low ranges. Correlationwas very good (r = 0.98) for WBC counts below 4.0×103/μL (4.0×109/L) but poor (r = 0.52) for PLT counts less than 50×103/μL (50×109/L). These findings prompted a program change in the analytical algorithm for PLT counts by Technicon Instruments Corp. After this change, an excellent correlation coefficient (r = 0.91) was demonstrated for PLT counts in the low range.

About this research paper

What this paper is about

The performance of the Technicon H-6000 was evaluated in the computerized, high-volume hematology laboratory at M.D. Anderson Hospital and Tumor Institute, Houston, Tex, and compared with that of reference instrumentation used for routine patient care. The within-run precision, linearity, and lack of carryover was excellent, and the day-to-day precision was satisfactory for the entire range of tested parameters (WBC, RBC, HGB, HCT, MCV, RDW, and PLT). Coefficients of correlation for these parameters were 0.98 or greater, with the exception of the MCV (r = 0.93). Special emphasis was placed on the correlation study of WBCs and PLTs at low ranges. Correlationwas very good (r = 0.98) for WBC counts below 4.0×103/μL (4.0×109/L) but poor (r = 0.52) for PLT counts less than 50×103/μL (50×109/L). These findings prompted a program change in the analytical algorithm for PLT counts by Technicon Instruments Corp. After this change, an excellent correlation coefficient (r = 0.91) was demonstrated for PLT counts in the low range.

Why it matters

OpenAlex reports 1 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

The performance of the Technicon H-6000 was evaluated in the computerized, high-volume hematology laboratory at M.D. Anderson Hospital and Tumor Institute, Houston, Tex, and compared with that of reference instrumentation used for routine patient care. The within-run precision, linearity, and lack of carryover was excellent, and the day-to-day precision was satisfactory for the entire range of tested parameters (WBC, RBC, HGB, HCT, MCV, RDW, and PLT). Coefficients of correlation for these parameters were 0.98 or greater, with the exception of the MCV (r = 0.93). Special emphasis was placed on the correlation study of WBCs and PLTs at low ranges. Correlationwas very good (r = 0.98) for WBC counts below 4.0×103/μL (4.0×109/L) but poor (r = 0.52) for PLT counts less than 50×103/μL (50×109/L). These findings prompted a program change in the analytical algorithm for PLT counts by Technicon Instruments Corp. After this change, an excellent correlation coefficient (r = 0.91) was demonstrated for PLT counts in the low range.

Key concepts: AutoAnalyzer, Hematology analyzer, Medicine, Hematology, Correlation coefficient, Coefficient of variation, Nuclear medicine, Internal medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
A Multiparameter Automated Hematology Analyzer: The Technicon H-6000 — Research Paper | ScholarLens