1999UCL Discovery (University College London)Requires access

Photon Counting with the Hamamatsu H6568 multi-anode photomultiplier

Soares, AJ, Ian Cullum, D.J. Miller, Gary Royle, R.D. Speller, J. C. Thomas

Open publisher page 0 citations

Abstract

We have evaluated the R5900-M16 multi-anode photomultiplier tube from Hamamatsu. Three tubes were tested for their uniformity of collection of photo-electrons by the dynode structure and for their uniformity of gain across the active area. Variations of less than 10% in two tubes and 30% in the third one were obtained. The cross-talk between closest neighbour pixels was found to be approximate to 1% (C) 2002 Elsevier Science B.V. All rights reserved.

About this research paper

What this paper is about

We have evaluated the R5900-M16 multi-anode photomultiplier tube from Hamamatsu. Three tubes were tested for their uniformity of collection of photo-electrons by the dynode structure and for their uniformity of gain across the active area. Variations of less than 10% in two tubes and 30% in the third one were obtained. The cross-talk between closest neighbour pixels was found to be approximate to 1% (C) 2002 Elsevier Science B.V. All rights reserved.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

We have evaluated the R5900-M16 multi-anode photomultiplier tube from Hamamatsu. Three tubes were tested for their uniformity of collection of photo-electrons by the dynode structure and for their uniformity of gain across the active area. Variations of less than 10% in two tubes and 30% in the third one were obtained. The cross-talk between closest neighbour pixels was found to be approximate to 1% (C) 2002 Elsevier Science B.V. All rights reserved.

Key concepts: Dynode, Photomultiplier, Optics, Anode, Photon counting, Physics, Photon, Electron