Comparison of the radiation dose from film and xeroradiography

R Evers, Michiel G. M. Roemer, D. Knappschafts-Krankenhaus

Open publisher page 0 citations

Abstract

Using a LiF dosimeter, comparative measurements were carried out on film mammography and xeromammography on surface and exit dose to the breast and scatter at the level of the gonads. Exit dose at 200 to 330 mR was approximately the same for both methods; incident dose at 6,800 mR for film mammography was significant higher than for xeromammography at 3,800 mR (Mo tube) or 1,200 mR (tungsten tube). Scatter at the level of the gonads for all three methods was between 4.4 and 12.8 mR. For xeroradiography of the knee joint, the radiation dose of 466 mR was approximately three times that for a film technique (160 mR). (orig.)

About this research paper

What this paper is about

Using a LiF dosimeter, comparative measurements were carried out on film mammography and xeromammography on surface and exit dose to the breast and scatter at the level of the gonads. Exit dose at 200 to 330 mR was approximately the same for both methods; incident dose at 6,800 mR for film mammography was significant higher than for xeromammography at 3,800 mR (Mo tube) or 1,200 mR (tungsten tube). Scatter at the level of the gonads for all three methods was between 4.4 and 12.8 mR. For xeroradiography of the knee joint, the radiation dose of 466 mR was approximately three times that for a film technique (160 mR). (orig.)

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Using a LiF dosimeter, comparative measurements were carried out on film mammography and xeromammography on surface and exit dose to the breast and scatter at the level of the gonads. Exit dose at 200 to 330 mR was approximately the same for both methods; incident dose at 6,800 mR for film mammography was significant higher than for xeromammography at 3,800 mR (Mo tube) or 1,200 mR (tungsten tube). Scatter at the level of the gonads for all three methods was between 4.4 and 12.8 mR. For xeroradiography of the knee joint, the radiation dose of 466 mR was approximately three times that for a film technique (160 mR). (orig.)

Key concepts: Xeroradiography, Nuclear medicine, Dosimeter, Radiation dose, Mammography, Dosimetry, Materials science, Medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
Comparison of the radiation dose from film and xeroradiography — Research Paper | ScholarLens