Clinical Application of Transcutaneous Oxygen Monitoring in Newborn Infants

Yong‐Kwei Tsau, Shen-Hui Chen

Open publisher page 0 citations

Abstract

The arterial oxygen tension was measured in 15 sick newborn infants using two different techniques simultaneously: (1) the conventional method using an arterial blood gas analyzer, and (2) transcutaneous oxygen monitoring using a surface electrode. Transcutaneous PO2 values (TcPO2) correlated well with the arterial values (PaO2). The calculated regression line was TcPO2=13.34+0.81 PaO2 with a correlation coefficient of r=0.86 (P<0.001). Transcutaneous oxygen tension measurements were also performed in 55 healthy newborn infants during normoxia (breathing room air) and hyperoxia (breathing 40% oxygen). A significant rise of TcPO2 was noted even during 40% oxygen inhalation indicating the importance of arterial PO2 monitoring in the newborns receiving oxygen therapy.

About this research paper

What this paper is about

The arterial oxygen tension was measured in 15 sick newborn infants using two different techniques simultaneously: (1) the conventional method using an arterial blood gas analyzer, and (2) transcutaneous oxygen monitoring using a surface electrode. Transcutaneous PO2 values (TcPO2) correlated well with the arterial values (PaO2). The calculated regression line was TcPO2=13.34+0.81 PaO2 with a correlation coefficient of r=0.86 (P<0.001). Transcutaneous oxygen tension measurements were also performed in 55 healthy newborn infants during normoxia (breathing room air) and hyperoxia (breathing 40% oxygen). A significant rise of TcPO2 was noted even during 40% oxygen inhalation indicating the importance of arterial PO2 monitoring in the newborns receiving oxygen therapy.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

The arterial oxygen tension was measured in 15 sick newborn infants using two different techniques simultaneously: (1) the conventional method using an arterial blood gas analyzer, and (2) transcutaneous oxygen monitoring using a surface electrode. Transcutaneous PO2 values (TcPO2) correlated well with the arterial values (PaO2). The calculated regression line was TcPO2=13.34+0.81 PaO2 with a correlation coefficient of r=0.86 (P<0.001). Transcutaneous oxygen tension measurements were also performed in 55 healthy newborn infants during normoxia (breathing room air) and hyperoxia (breathing 40% oxygen). A significant rise of TcPO2 was noted even during 40% oxygen inhalation indicating the importance of arterial PO2 monitoring in the newborns receiving oxygen therapy.

Key concepts: Medicine, Arterial oxygen tension, Hyperoxia, Oxygen, Arterial blood, Breathing, Breathing gas, Anesthesia